6-3-4 transducer block do202, 6-3-4 transducer block do202 -41 – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 46

Advertising
background image

IM 12A00A01-61E

Foundation Fieldbus 3-41

Index Parameter

Factory

valid range

description

Name

default

2000

BLK_DATA

TAG: “TB”

General information about the function block

2001

ST_REV

-

-

The revision level of the static data associated with the

function block. The revision value will be incremented

each time a static parameter value in the block is

changed.

2002

TAG_DESC

“ “

The user description of the intended application of the

block.

2003

STRATEGY

1

The strategy field can be used to identify grouping of

blocks. This data is not checked or processed by the

block

2004

ALERT_KEY

1

The identification number of the plant unit. This

information may be used in the host for sorting alarms,

etc.

2005

MODE_BLK

AUTO

The actual, target, permitted, and normal modes of the

block

2006

BLOCK_ERR

-

This parameter reflects the error status associated with a

block. It is a bit-string, so that multiple errors can be shown.

2007

UPDATE_EVT

-

The alert is generated by any change to the static data.

2008

BLOCK_ALM

-

The block alarm is used for all configuration error,

hardware connection failure or system problems in the

block. The cause of the alert is entered in the subcode

field. The first alert to become active will set Active

status in Status attribute.

2009

TRANSDUCER

A directory that specifies the number and starting

_DIRECTORY

indices of the transducers.

2010

TRANSDUCER_TYPE

DO

DO

Yokogawa specific Dissolved Oxygen transducer block

2011

XD_ERROR

-

The error code in transducer:

- No failure

- Electronics failure

- I/O failure

- Mechanical failure

2012

COLLECTION

-

A directory that specifies the number, starting indices

_DIRECTORY

and DD item IDs of the data collection in each

transducer within a transducer block.

2013

PRIMARY_VALUE

Dissolved Oxygen

Dissolved Oxygen

The type of measurement represented by primary

_TYPE

value.

2014

PRIMARY_VALUE

-

0 to 50 ppm

Primary measuring value of the instrument including

0 to 1999 ppb

status information.

0 to 600 %

2015

PRIMARY_VALUE

-

0 to 50 ppm

The measuring range of the instrument associated with

_RANGE

0 to 1999 ppb

the selected unit type.

0 to 600 %

2016

PRIMARY_VALUE_UNIT ppm

ppm, ppb, %

Select the unit in which the primary value must be

presented

2017

SENSOR_TYPE

Polarographic

Polarographic, Galvanic Select the type of sensor connected to the instrument,

_OXYGEN

Galvanic or Polarographic

2018

SAMPLE_CAL

-

0 to 50 ppm

Enter the correct (current) measured value. This value

0 to 1999 ppb

is used to calculate the new sensitivity (slope)

0 to 600 %

2019

ZERO_CURRENT

0

+- ZERO_CURRENT

Directly overwrites the current zero current, which is

_LIMIT (param 2037)

the leak current of the sensor

2020

SENSITIVITY

7.5

1.000 to 1999 nA/ppm

Directly overwrites the sensitivity (slope) of the sensor

2021

AMP_STABILIZE_TIME 60

5 to 600 seconds

Stability criteria used during automatic calibration.

2022

AMP_SPAN_STABILIZE 0.1

0 to 50 ppm

Stability criteria used during automatic span calibration.

_VALUE

3-6-3-4 Transducer Block dO202