3-5-2 function block execution control, 3-6 block setting, 3-6-1 link object – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 17: 3-5-2 function block execution control -12, 3-6 block setting -12, 3-6-1 link object -12, Set link objects as shown in table 3.8

Advertising
background image

IM 12A00A01-61E

3-12 Foundation Fieldbus

Table 3.6� VCR List

VCR List

Index (SM)

VCR Number

Factory Setting

293

1

For system management (Fixed)

294

2

Server (LocalAddr = 0xF3)

295

3

Server (LocalAddr = 0xF4)

296

4

Server (LocalAddr = 0xF7)

297

5

Trend Source (LocalAddr = 0x07, Remote Address=0x111)

298

6

Publisher for AI1 (LocalAddr = 0x20)

299

7

Alert Source (LocalAddr = 0x07, Remote Address=0x110)

300

8

Server (LocalAddr = 0xF9)

301 to 325

to 325

9 to 33

Not set

3-3-5-2 Function Block Execution Control
According to the instructions given in Section 3-3-3, set the execution cycle of the function blocks and
schedule of execution.

3-3-6� Block Setting

Set the parameter for function block VFD.

3-3-6�-1 Link Object
Link object combines the data voluntarily sent by the function block with VCR. The EXA has 40 link objects.
A single link object specifies one combination. Each link object has the parameters listed in Table 3.7.
Parameters must be changed together for each VCR because the modifications made to each parameter
may cause inconsistent operation.

Table 3.7 Link Object Parameters

Sub-

index

Parameters

Description

1

LocalIndex

Sets the index of function block parameters to be combined; set “0” for Trend and Alert.

2

VcrNumber

Sets the index of VCR to be combined. If set to “0”, this link object is not used.

3

RemoteIndex

Not used in EXA. Set to “0”.

4

ServiceOperation

Set one of the following. Only one link object is used for Alert and/or Trend.
0: Undefined
2: Publisher
6: Alert
7: Trend

5

StaleCountLimit

Set the maximum number of consecutive stale input values which may be received before the input
status is set to BAD. To avoid the unnecessary mode transition caused when the data is not correctly
received by subscriber, set this parameter to “2” or more.

Set link objects as shown in Table 3.8.

Table 3.8 Factory-Settings of Link Objects (example)

Index

Link Object#

Factory Settings

30000

1

AI1.OUT

VCR#6

30001

2

Trend

VCR#5

30002

3

Alert

VCR#7

30003 to 30039

4 to 40

Not used