Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 41

Advertising
background image

IM 12A00A01-61E

3-36 Foundation Fieldbus

Index Parameter name

Factory

Valid

Description

Default

Range

2034

REFERENCE_

25

0 to 100 ºC,

Conductivity can be process compensated to a standard

TEMPERATURE

32 to 212 ºF

reference temperature. Mostly 20ºC or 25ºC is used

2035

COMP_METHOD

NaCl

NaCl, TC, matrix

Method of process temp. compensation for the primary value

2036

COMP_MATRIX_SEL

HCl

HCl cation (0-80 °C)

When matrix compensation is required one can make a

Ammonia, (0-80 °C)

selection out of 5 predefined matrices and one user

Ammonia, (0-80 °C)

definable matrix

HCl (0-5%, 0-60 °C)

NaOH (0-5%, 0-100 °C),

User defined

2037

TERTIARY_COMP_METHOD NaCl

NaCl, TC, matrix

Method of process temperature compensation for the

second conductivity value

2038

TERT_TEMPERATURE

2.1

0 to 3.5%/ºC

Process temperature compensation factor for the

_COEFF

second conductivity value

2039

ALARM_SUM

-

2040

DEV_ALARM

-

Device Alarm is used to give the status of the

analyser. See separate table for error messages

2041

LOGBOOK1_RESET

Idle

Idle, Reset

Reset the pointer to the first (oldest) event in logbook 1

2042

LOGBOOK1_EVENT

-

Event whereto the pointer is referenced. When parameter is

read, the pointer is increased by one.

2043

LOGBOOK2_RESET

Idle

Idle, Reset

Reset the pointer to the first (oldest) event in logbook 2

2044

LOGBOOK2_EVENT

-

Event whereto the pointer is referenced. When parameter is

read, the pointer is increased by one

2045

LOGBOOK_CONFIG[16]

-

Per event one can decide whether it should be logged and

in which logbook (1 or 2) it should be logged

2046

TEST_1

-

....

-

2047 to 2057 are, like 2046 and 2058, service parameters

2058

TEST_13

-

2059

CALIB_SENSOR_CONST

0.1

0.005~50

The calibrated cell constant [1/cm].

2060

MATRIX_TEMP_RANGE

-20~250, 0~500

The temperature values of the user defined matrix.

2061 SOLUTION_1

0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 1.

2062 SOLUTION_2

0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 2.

2063 SOLUTION_3

0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 3.

2064 SOLUTION_4

0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 4.

2065 SOLUTION_5

0~1.999S/cm, 0~999MΩ•cm The conductivity/resistivity values of solution 5.

2066

CONCENTRATION_

disabled

0=Not initialized,

Enabling/Disabling of concentration measurement.

MEASUREMENT

1=disabled, 2=enabled

2067

CONCENTRATION_0

0

0~100

The concentration [%] corresponding to the 0 percent table value

2068

CONCENTRATION_100

100

0~100

The concentration [%] corresponding to the 100 percent table value

2069 CONC._TABLE_LOW

0~1.999S/cm, 0~999MΩ•cm The first part of the concentration table (0 till 30 percent).

2070 CONC._TABLE_MID

0~1.999S/cm, 0~999MΩ•cm The second part of the concentration table (35 till 65 percent).

2071 CONC._TABLE_HIGH

0~1.999S/cm, 0~999MΩ•cm The third part of the concentration table (70 till 100 percent).

2072 E5_LIMIT

0.25, 4

0~0.5S, 0~10MΩ

The upper range accuracy limit of the analyzer [S or Ohm].

2073

E6_LIMIT

1

μ, 1M

0~0.5S, 0~10MΩ

The lower range accuracy limit of the analyzer [S or Ohm].

2074

DISPLAY_RESOLUTION

1

0 to 8

Display resolution, 0, Not initialized (can not change)

1,Auto Ranging, 2, x.xxx

μS/cm, MΩ•cm,

3, xx.xx

μS/cm, MΩ•cm, 4, xxx.x μS/cm, MΩ•cm,

5, x.xxx mS/cm, kΩ•cm, 6, xx.xx mS/cm, kΩ•cm,
7, xxx.x mS/cm, kΩ•cm, 8, xxxx mS/cm, kΩ•cm

2075

PASSCODE_MAINT

0

0, 111, 333, 777, 888,

Passcode used to protect the maintenance menu

123, 957, 331, 546, 847

2076

PASSCODE_COMM

0

0, 111, 333, 777, 888,

Passcode used to protect the commissioning menu.

123, 957, 331, 546, 847

2077

PASSCODE_SERVICE

0

0, 111, 333, 777, 888,

Passcode used to protect the service menu.

123, 957, 331, 546, 847

2078

ERROR_CONFIG*

0=soft, 1=hard

Soft/Hard fail configuration (checked = hard unchecked = soft).

2079

CONFIGURATION*

polarization on,

Instrument specific configuration. See below for correct

USP off,

bitnr. setting autoreturn on

2080

TRANSMITTER_TIME

Transmitter time.”

2081

TEST_14

Service parameter.