Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 39

Advertising
background image

IM 12A00A01-61E

3-34 Foundation Fieldbus

Index Parameter name

Factory

Valid

Description

Default

Range

2075 PASSCODE_MAINTENANCE 0

0, 111, 333, 777,

Passcode used to protect the maintenance menu.

888, 123, 957,

331, 546, 847

2076 PASSCODE_COMMISSIONING 0

0, 111, 333, 777,

Passcode used to protect the commissioning menu.

888, 123, 957,

331, 546, 847

2077 PASSCODE_SERVICE

0

0, 111, 333, 777,

Passcode used to protect the service menu.

888, 123, 957,

331, 546, 847

2078 SAMPLE_PV

Process value of the sample taken.

2079 SAMPLE_PV2

Second process value of the sample taken.

2080 SAMPLE_TEMP

Temperature of the sample taken.

2081 ERROR_CONFIG*

Soft/Hard fail configuration

(checked = hard; unchecked = soft)

2082 CONFIGURATION*

Instrument specific configuration

2083 TRANSMITTER_TIME

Transmitter time.

2084 TEST_14

Service parameter.

* Enumerated parameters (4 bytes, 32 bits) where each bit can be set individually.

Error_config
0x00002000 (bit 14), “input 1 impedance exceeds low limit (E4.1)” ,
0x00001000 (bit 13), “input 1 impedance exceeds high limit (E5.1)” ,
0x00000800 (bit 12), “input 2 impedance exceeds low limit (E4.2)” ,
0x00000400 (bit 11) “input 2 impedance exceeds high limit (E5.2)” ,
0x00000200 (bit 10), “temperature sensor open (E7)” ,
0x00000100 (bit 9), “temperature sensor shorted (E8)” ,
0x00800000 (bit 24), “primary value exceeds limits (E9)” ,

0x00010000 (bit 17), “calibration timer expired (E16)”

Configuration
0x08000000 (bit 28), “Process compensation enabled” ,
0x00800000 (bit 24), “Aspot check enabled” ,
0x00400000 (bit 23), “Slope check enabled” ,
0x00200000 (bit 22), “Display resolution 0.01 pH” ,
0x00100000 (bit 21), “Auto return enabled” ,
0x00080000 (bit 20), “Maintenance timer enabled” ,
0x00002000 (bit 14), “Input 1 impedance measurement enabled” ,
0x00008000 (bit 16), “Input 1 impedance high” ,
0x00004000 (bit 15), “Input 1 impedance compensation enabled” ,
0x00000400 (bit 11), “Input 2 impedance measurement enabled” ,
0x00001000 (bit 13), “Input 2 impedance high” ,
0x00000800 (bit 12), “Input 2 impedance compensation enabled” ,
0x00000010 (bit 5), “Take sample”