Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 73

Advertising
background image

Profibus 4-15

IM 12A00A01-61E

4-4-3-3. Function Block Parameters ISC202 (continued)

Parameter

default

Alternatives

Unit r/W Slot, Index data Type

(byte,bit) (bytes)

Concentration 0%

0

0~100

%

R/W 4, 121

float (4)

Concentration 100%

100

0~100

%

R/W 4, 122

float (4)

Concentration Table 0%

0

0~1.999

S/cm R/W 4, 123

float (4)

Concentration Table 5%

0.05

S/cm R/W 4, 124

float (4)

Concentration Table 10% 0.1

S/cm R/W 4, 125

float (4)

Concentration Table 15% 0.15

S/cm R/W 4, 126

float (4)

Concentration Table 20% 0.2

S/cm R/W 4, 127

float (4)

Concentration Table 25% 0.25

S/cm R/W 4, 128

float (4)

Concentration Table 30% 0.3

S/cm R/W 4, 129

float (4)

Concentration Table 35% 0.35

0~1.999

S/cm R/W 4, 130

float (4)

Concentration Table 40% 0.4

S/cm R/W 4, 131

float (4)

Concentration Table 45% 0.45

S/cm R/W 4, 132

float (4)

Concentration Table 50% 0.5

S/cm R/W 4, 133

float (4)

Concentration Table 55% 0.55

S/cm R/W 4, 134

float (4)

Concentration Table 60% 0.6

S/cm R/W 4, 135

float (4)

Concentration Table 65% 0.65

S/cm R/W 4, 136

float (4)

Concentration Table 70% 0.7

0~1.999

S/cm R/W 4, 137

float (4)

Concentration Table 75% 0.75

S/cm R/W 4, 138

float (4)

Concentration Table 80% 0.8

S/cm R/W 4, 139

float (4)

Concentration Table 85% 0.85

S/cm R/W 4, 140

float (4)

Concentration Table 90% 0.9

S/cm R/W 4, 141

float (4)

Concentration Table 95% 0.95

S/cm R/W 4, 142

float (4)

Concentration Table 100% 1

S/cm R/W 4, 143

float (4)

Calibration Parameters

Nominal cell constant

1.88

0.005~50

1/cm R/W 4, 30

float (4)

Calibrated cell constant

0.1

0.005~50

1/cm R

4, 89

float (4)

Calibration method

Not calibrated(0)

1 point(107), 2 point(108)

R/W 4, 34

unsigned8 (1)

Diagnostic Settings

E5 Limit

3

0~0.5Ω

S

R/W 4, 144

float (4)

E6 Limit

0.000005

0~0.5Ω

S

R/W 4, 145

float (4)

Conductivity exceeds

high limit (E5)

Hard fail(1)

Soft fail(0)

4, 150(2.4)

Conductivity exceeds

low limit (E6)

Hard fail(1)

Soft fail(0)

4, 150(2.5)

Temp. sensor open (E7) Hard fail(1)

Soft fail(0)

4, 150(2.6)

Temp. sensor shorted (E8) Hard fail(1)

Soft fail(0)

4, 150(2.7)

Passcode Configuration

Passcode Maintenance

0

0, 111, 333, 777, 888, 123, 957, 331, 546, 847

R/W 4, 147

unsigned16 (2)

Passcode Commissioning 0

0, 111, 333, 777, 888, 123, 957, 331, 546, 847

R/W 4, 148

unsigned16 (2)

Passcode Service

0

0, 111, 333, 777, 888, 123, 957, 331, 546, 847

R/W 4, 149

unsigned16 (2)

display Settings

Display Resolution

Auto Ranging(1)

auto(1), xxx.x µS/cm(4) x.xxx mS/cm(5), xx.xx

mS/cm(6) xxx.x mS/cm(7), xxxx mS/cm(8)

R

4, 146

unsigned8 (1)

Auto return

Enabled(1)

Disable(0)

R

4, 151(1.0)

Logbook Configuration

Power up

Logbook 2(3)

off(1), logbook1(2), logbook2(3)

R/W 4, 59

unsigned8 (1)

Power down

Logbook 2(3)

R/W 4, 60

unsigned8 (1)

System error

Logbook 1(2)

R/W 4, 61

unsigned8 (1)

Defaults loaded

Not logged(1)

R/W 4, 62

unsigned8 (1)

Logbook erased

Not logged(1)

R/W 4, 63

unsigned8 (1)

Init performed

Logbook 1(2)

R/W 4, 64

unsigned8 (1)

Error on

Not logged(1)

R/W 4, 65

unsigned8 (1)

Error off

Not logged(1)

R/W 4, 66

unsigned8 (1)

Temperature adjust

Logbook 1(2)

R/W 4, 67

unsigned8 (1)

Cell constant

Logbook 1(2)

R/W 4, 68

unsigned8 (1)

Air calibration

Logbook 1(2)

R/W 4, 69

unsigned8 (1)

Advertising