3-42 Foundation Fieldbus

Index Parameter

Factory

valid range

description

Name

default

2023

AMP_ZERO_STABILIZE -

0 to 50 ppm

Stability criteria used during automatic zero calibration.

_VALUE

0 to 1999 ppb

0 to 600 %

2024

SALINITY

0

0 to 99.9 ppt

Salinity value of the process liquid that can be used to

compensate the DO value

2025

BAR_PRESSURE

101.3

0.0 to 999 kPa

Manual value of the pressure used to compensate the

DO reading (during calibration)

2026

BAR_PRESSURE_UNIT kPa

kPa

The engineering unit of barometric pressure value

2027

PERCENT_SATURATION 101.3

0.0 to 999 kPa

Manual value of the pressure used to compensate

_PRESSURE

the DO reading (during measuring)

2028

CHLORINE_CALI-

not used

BRATION_RANGES

2029

SECONDARY_VALUE

-

-20 to 150 °C, -4

Temperature value

to 302 °F

2030

SECONDARY_VALUE

°C

°C, °F

Temperature unit

_UNIT

2031

SENSOR_TEMP_COMP automatic

automatic, manual

Select manual when no temperature element is available

and the temperature is stable and select auto when a

temperature element is available

2032

SENSOR_TEMP_MAN

25°

0 - 100 °C 32 - 212 °F

manual temperature value

_VALUE

2033

SENSOR_TYPE_TEMP NTC22K

NTC22K, Pb36, PT1000

Temperature element used

2034

TEMP_SENSOR_CAL

-

-20 to 150 °C, -4

Enter the correct (current) measured value. This value

to 302 °F

is used to calculate the temperature offset

2035

SENSOR_CURRENT

-

0 to 1200 nA,

The actual cell current of the DO sensor

0 to 50 µA

2036

PERCENT_SATURATION -

0 to 600 %

The DO value expressed as a percentage of the

maximum saturation.

2037

ZERO_CURRENT_LIMIT

0 to 199.9 nA,

Limit used during zero calibration for the sensor

0 to 19.99 µA

current (different for polarographic or galvanic sensor)

2038

ZERO_CAL

-

0 to 50 ppm

Enter the correct (current) measured value. This value

0 to 1999 ppb

is used to calculate the new offset (zero)

0 to 600 %

2039

RESERVED1

not used

2040

RESERVED2

not used

2041

RESERVED3

not used

2042

TRANSMITTER_TIME[6]

Transmitter time.

2043

PASSCODE

0

0, 111, 333, 777, 888, Passcode used to protect the maintenance menu.

_MAINTENANCE

123, 957, 331, 546, 847

2044

PASSCODE

0

0, 111, 333, 777, 888, Passcode used to protect the commissioning menu

_COMMISSIONING

123, 957, 331, 546, 847

2045

PASSCODE_SERVICE

0

0, 111, 333, 777, 888, Passcode used to protect the service menu.

123, 957, 331, 546, 847

2046

LOGBOOK1_RESET

-

Reset the pointer to the first (oldest) event in logbook 1.

2047

LOGBOOK1_EVENT

-

Event whereto the pointer is referenced. When

parameter is read, the pointer is increased by one.

2048

LOGBOOK2_RESET

-

Reset the pointer to the first (oldest) event in logbook 2.

2049

LOGBOOK2_EVENT

-

Event whereto the pointer is referenced. When

parameter is read, the pointer is increased by one.

2050

LOGBOOK_CONFIG[25]

-

Per event one can decide whether it should be logged

and in which logbook (1 or 2) it should be logged.

2051

CALL_MAINT_TIME

250

1 to 250 days

The interval of the maintenance timer

_RELOAD

2052

CALL_MAINT_TIME

250

1 to 250 days

The remaining days till maintenance is required.

_COUNTDOWN

2053

ERROR_CONFIG*

E9,E2,E7,E8 hard fail.

Soft/Hard fail configuration (checked = hard;

E16 soft fail

unchecked = soft)