3-30 Foundation Fieldbus

Relative Parameter

Factory

Write

Index

Name

Default

Mode

Explanation

0

Block Header

TAG: “AI1” or

Block Tag Information on this block such as Block Tag, DD

“AI2” or “AI3”

= O/S

Revision, Execution Time etc.

1

ST_REV

The revision level of the static data associated with the function block. The

revision value will be incremented each time a static parameter value in the block

is changed.

2

TAG_DESC

(blank)

AUTO

The user description of the intended application of the block.

3

STRATEGY

1

AUTO

The strategy field can be used to identify grouping of blocks. This data is not

checked or processed by the block.

4

ALERT_KEY

1

AUTO

The identification number of the plant unit. This information may be used in the

host for sorting alarms, etc.

5

MODE_BLK

AUTO

AUTO

The actual, target, permitted, and normal modes of the block.

6

BLOCK_ERR

This parameter reflects the error status associated with the hardware or software

components associated with a block. It is a bit string, so that multiple errors may

be shown.

7

PV

Either the primary analog value for use in executing the function, or a process

value associated with it. May also be calculated from the READBACK value of an

AO block.

8

OUT

Value=

The primary analog value calculated as a result of

MAN

executing the function.

9

SIMULATE

Disable

AUTO

Allows the transducer analog input or output to the block to be manually supplied

when simulate is enabled. When simulation is disabled, the simulate value and

status track the actual value and status.

10

XD_SCALE

O/S

The high and low scale values, engineering units

code, and number of digits to the right of the decimal point used with the value

obtained from the transducer for a specified channel. Refer to Table 3.18 to 3.21

for the unit available.

11

OUT_SCALE

O/S

The high and low scale values, engineering units

code, and number of digits to the right of the decimal point to be used in

displaying the OUT parameter and parameters which have the same scaling as

OUT.

12

GRANT_DENY

0

AUTO

Options for controlling access of host computers and local control panels to

operating, tuning and alarm parameters of the block.

13

IO_OPTS

0

O/S

Options which the user may select to alter input and output block processing

bit6: Low cutoff

14

STATUS_OPTS Propagate Fault

O/S

Options which the user may select in the block

Forward

processing of status

15

CHANNEL

AI1: 1

O/S

The number of the logical hardware channel that is

AI2: 2

connected to this I/O block. This information defines the transducer to be

AI3: 3

used going to or from the physical world.

16

L_TYPE

Direct (1)

MAN

Deterines if the values passed by the transducer time of order block to the AI

block may be used directly (Direct) or if the value is in different units and must be

converted linearly (Indirect), or with square root (Ind Sqr Root), using the input

range defined by the transducer and the associated output range.

17

LOW_CUT

Linear: 0%

AUTO

Sets low cut point of output. This low cut value become available by setting

“Low cutoff“ to “IO_OPTS“.

18

PV_FTIME

2sec

AUTO

Time constant of a single exponential filter for the PV, in seconds.

19

FIELD_VAL

Raw value of the field device in percent of thePV range, with a status reflecting

the Transducer condition, before signal characterization (L_TYPE) or filtering

(PV_FTIME).

20

UPDATE_EVT

This alert is generated by any change to the static data.

21

BLOCK_ALM

The block alarm is used for all configuration, hardware, connection failure or

system problems in the block. The cause of the alert is entered in the subcode

field. The first alert to become active will set the Active status in the Status

attribute. As soon as the Unreported status is cleared by the alert reporting task,

another block alert may be reported without clearing the Active status, if the

subcode has changed.