3-26 Foundation Fieldbus

DEVICE_STATUS_2 PH202

Hexadecimal Display through DD

0x80000000

0x40000000

0x20000000

0x10000000

0x08000000

0x04000000

0x02000000

0x01000000

0x00800000

0x00400000

0x00200000

0x00100000

0x00080000 FF interface checksum error

0x00040000 EXA checksum error (E21)

0x00020000 Internal communication failure

0x00010000 FF interface eeprom failure

0x00008000 EXA eeprom failure (E20)

0x00004000 mismatch between FF- and EXA parameter

0x00002000

0x00001000

0x00000800

0x00000400

0x00000200

0x00000100 calibration timer expired (E16)

0x00000080 reference impedance exceeds low limit (E4.2)

0x00000040 reference impedance exceeds high limit (E5.2)

0x00000020 glass impedance exceeds low limit (E4.1)

0x00000010 glass impedance exceeds high limit (E5.1)

0x00000008 temperature sensor shorted (E8)

0x00000004 temperature sensor open (E7)

0x00000002 tertiary value exceeds limits (E12)

0x00000001 primary value exceeds limits (E9)

DEVICE_STATUS_2 SC202

Hexadecimal Display through DD

0x80000000

0x40000000

0x20000000

0x10000000

0x08000000

0x04000000

0x02000000

0x01000000

0x00800000

0x00400000

0x00200000

0x00100000

0x00080000 FF interface checksum error

0x00040000 EXA checksum error (E21)

0x00020000 Hart communication failure

0x00010000 FF interface eeprom failure

0x00008000 EXA eeprom failure (E20)

0x00004000 mismatch between FF- and EXA parameter

0x00002000

0x00001000

0x00000800

0x00000400

0x00000200

0x00000100 matrix error (E4)

0x00000080 concentration table error (E18)

0x00000040 conductivity exceeds usp limit (E13)

0x00000020 polarization detected (E1)

0x00000010 temperature compensation error (E2)

0x00000008 temperature sensor shorted (E8)

0x00000004 temperature sensor open (E7)

0x00000002 conductivity exceeds low limit (E6)

0x00000001 conductivity exceeds high limit (E5)