2-2 host setting, 2-3 bus power on, 2-2 host setting -4 – Yokogawa DO202 2-Wire Dissolved Oxygen Analyzer User Manual

Page 9: 2-3 bus power on -4, Important

Advertising
background image

IM 12A00A01-61E

3-4 Foundation Fieldbus

3-2-2 Host Setting

To activate Fieldbus, the following settings are required for the host.

IMPORTANT

Do not turn off the power immediately after setting. When the parameters are saved to the EEPROM, the

redundant processing is executed for an improvement of reliability. If the power is turned off within 60 seconds after

setting is made, the modified parameters are not saved and the settings may return to the original values.

Table 3.1 Operation Parameters

Symbol

Parameter

Description and Settings

V (ST)

Slot-Time

Set 4 or greater value.

V (MID)

Minimum-Inter-PDU-Delay

Set 4 or greater value.

V (MRD) Maximum-Reply-Delay

Set so that V (MRD) X V (ST) is 12 or greater

V (FUN)

First-Unpolled-Node

Define the first address that can be used by the host. Set 0x15 or greater.

V (NUN)

Number-of-consecutive-Unpolled-Node

This sets the number of consecutive unpolled nodes. EXA address is

factory-set to 0xEB. Set this address to be within the range of the BASIC

device in Figure 3.5.

Not used

LM device

Not used

Basic device

Default address

Portable-device address

V (FUN)

V (FUN) + V (NUN)

V (NUN)

0xFF

0xFC

0xFB

0xF8

0xF7

0x00

0x0F

0x10
0x13

0x14

Bridge device

Note 1: LM device: with bus control function (Link Master function)
Note 2: BASIC device: without bus control function

EXA

Figure 3.5 Available Address Range

3-2-3 Bus Power ON

Turn on the power of the host and the bus. First all segments of the display are lit, then the display begins

to operate. If the indicator is not lit, check the polarity of the power supply.

Using the host device display function, check that the EXA is in operation on the bus.

Unless otherwise specified, the following settings are in effect when shipped from the factory.

PH202

SC202

ISC202

DO202

PD tag

PH1001

SC1001

ISC1001

DO1001

Node addr.

232

233

234

235

DEV_TYPE

0x0830

0x0831

0x0832

0x0833

If no EXA is detected, check the available address range and the polarity of the power supply. If the node

address and PD tag are not specified when ordering, default value is factory set. If two or more EXA’s are

connected at a time with default value, only one EXA will be detected from the host as EXA’s have the

same initial address.

Separately connect each EXA and set a different address for each.