3-20 Foundation Fieldbus

Table 3.21 Unit Index by XD_SCALE DO202

Channel

FF parameters

Service code

XD_SCALE.UNITS

2

FF2030

SC11 (set to 0)

°C (1001)

2

FF2030

SC11 (set to 1)

°F (1002)

1

FF2016

SC56 (set to 0)

ppm (1423)

1

FF2016

SC56 (set to 1)

ppb (1424)

1

FF2016

SC56 (set to 2)

% (1342)

3

-

Default %

% (1342)

4

-

Default nA

nA (1213)

L_TYPE:
Specifies the operation function of the AI block. If set to “Direct”, the input delivered to CHANNEL is directly
reflected on OUT. If set to “Indirect”, scaling by XD_SCALE and OUT_SCALE is carried out and is reflected
on OUT. If set to “Indirect SQRT”, after scaling by XD_SCALE, the square root is extracted and the value
scaled by OUT_SCALE is reflected on OUT.

Example:
Channel range is defined as 0 to 100 °C but °F units is required for HOST display.
Set the following parameters:

XD_SCALE:

EU@0% = 0 °C

EU@100% = 100 °C

Unit = °C

Decimal point = 2

OUT_SCALE: EU@0% = 32 °F

EU@100% = 212 °F

Unit = °F

Decimal point = 2

0%, 0ºC

100%,

212ºF

0%, 32ºF

CHANNEL

PV / OUT

OUT_SCALE

XD_SCALE

100%, 100ºC

Figure 3.11 Scaling applied to temperature conversion.

PV_FTIME:
Sets the time constant of the damping function within AI block (primary delay) in seconds.

Alarm Priority:
Indicates the priority of the process alarm. If a value of 3 or greater is set, an alarm is transmitted.
The factory default is 0.
Four types of alarm can be set:
HI_PRI, HI_HI_PRI, LO_PRI, and LO_LO_PRI.

Alarm Threshold:
Sets the threshold at which a process alarm is generated. The factory default setting is a value that does
not generate an alarm.
Four types of alarm can be set:
HI_LIM, HI_HI_LIM, LO_LIM, and LO_LO_LIM.