Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 11

Advertising
background image

3

IM 12B07C05-01E

2

GENERAL

SPECIFICATIONS

F) Contact input

: Remote wash cycle start.

G) Temperature compensation

Function

: Automatic or manual

Compensation to Nernst equation

Process compensation by configurable temperature coefficient,

NEN6411 for water or strong acids/bases or programmable matrix

H) Calibration

: Semi-automatic 1 or 2 point calibration using pre-configured NIST,

US, DIN buffer tables 4, 7 & 9, or with user defined buffer tables, with

automatic stability check

Manual adjustment to grab sample

I) Logbook

: Software record of important events and diagnostic data readily

available in the display or through HART®.

J) Display

: Graphical Quarter VGA (320 x 240 pixels) LCD with LED backlight and

touchscreen. Plain language messages in English, German, French,

Spanish, Italian, Swedish ,Portuguese and Japanese.

K) Shipping details

Package size

: 290 x 300 x 290 mm (L x W x D) (11.5 x 11.8 x 11.5 inch)

Package weight

: app. 2.5 kg (5.5 lbs)

Converter weight : app. 1.5 kg

L) Housing

: Cast aluminum housing with chemically resistant coating;

Polycarbonate cover with Polycarbonate flexible window

: Protection IP66 / NEMA 4X / CSA Type 3S

Colour

: Silver grey

PH450G-A(D)-A

: IP66 cable glands are supplied with the unit

PH450G-A(D)-U

: NEMA 4X close up plugs are mounted in the unused cable entry holes

and can be replaced by conduit fittings as required

Pipe, Panel or Wall mounting using optional hardware

Optional conduit adapter

G1/2, 1/2NPT or M20 female

M) Power supply

PH450G-A

: Ratings; 100-240 V AC Acceptable range; 90 to 264 V AC

Ratings; 50/60 Hz

Acceptable range; 50 Hz ±5%, 60 Hz ±5%

Power Consumption; 15 VA

PH450G-D

: Ratings; 12-24 V DC

Acceptable range; 10.8 to 26.4 V DC

Power Consumption; 10 W

N) Safety and EMC conforming standards

,

Safety

: EN 61010-1

CSA C22.2 No.61010-1

UL 61010-1

FM3611 Class I, Div.2, Group ABCD,T6 for Ta -20 to 55ºC

EMC

: EN 61326-1 Class A, Table 2 (For use in industrial locations)

EN 61326-2-3

EN 61000-3-2 Class A

EN 61000-3-3

Korea Electromagnetic Conformity Standard

Installation altitude: 2000 m or less

Category based on IEC 61010: II (Note)

Pollution degree based on IEC 61010: 2 (Note)

Note: Installation category, called over-voltage category, specifies impulse withstand voltage.

Category II is for electrical equipment.

Pollution degree indicates the degree of existence of solid, liquid, gas or other

inclusions which may reduce dielectric strength. Degree 2 is the normal indoor

environment.