47

IM 12B07C05-01E

9

QUALITY

INSPECTION

2/3

QIS 12B07C05-01E

This test is done on the “HIF” display of “Factory Mode”.

a. Touch the [Setup] icon.

b. Touch the [Commissioning].

c. Touch the [Advanced setup].

d. Touch the [Factory adjustment].

e. Enter a password.

f. Select the [Factory Mode] in “Key.”

g. Select the [HIF] in “Execute.”

Select the [Input1(pH)] in “mV input.” When the standard voltage source to the

corresponding value of “Simulation input (mV)” in Table 1 is set, check the data display

and the value must be within the range shown in Table 1.

Table 1

Simulation input (mV)

Check Point (pH)

Data Display (pH)

414.1

0

0 ±0.01

0

7

7 ±0.01

-414.1

14

14 ±0.01

After the above test is completed, select the [Input1(ORP)] in “mV input.”

When the standard voltage source to the corresponding value of “Simulation input (mV)”

in Table 2 is set, check the data display and the value must be within the range shown in

Table 2.

Table 2

Simulation input (mV)

Check Point (ORP)

Data Display (mV)

-1500

-1500

-1500 ±1

0

0

0

±1

1500

1500

1500 ±1

After the above test is completed, touch the [Exit] to return to the “HIF” display.

3.4 Temperature

Indication Check

Following Section 3.3, select the [PT1000] in “Temperature” of the “HIF” display.

In this state, change the resistance of the decade resistance box and check the data

display. The value on the data display must be within the range shown in Table 3.

Table 3

Temperature (°C)

Resistance (Ω) of

Decade Resistance Box

Data Display (°C)

–10

960.9

–10 ±0.3

25

1097.3

25 ±0.3

120

1460.6

120 ±0.3

After the above test is completed, touch the [Exit] to return to the “HIF” display.

3.5 Current Output Test

Following Section 3.4, select the [Check] in “mA outputs” of the “HIF” display. “Set value

4.000 mA” appears at the bottom of the display. Select “Next value” in the “Command”

and touch “Enter,” the value on the data display increases in steps of 4 mA. Check the

current outputs 1 and 2 corresponding to the data display, the current output must be

within the range shown in Table 4.