3-7. Sensor wiring

Refer to figure 3-10, which includes drawings

that outline sensor wiring.

The EXAxt 450 can be used with a wide range

of commercially available sensor types, both

from Yokogawa and other manufacturers. The

sensor systems from Yokogawa fall into two

categories; the ones that use a fixed cable and

the ones with separate cables.

To connect sensors with fixed cables, simply

match the terminal numbers in the instrument with

the identification numbers on the cable ends.

The separate sensors and cables are not num-

bered, but instead use a color-coding system.

The electrodes have a colored band incorpo-

rated in the label on the connection cap:

• Red for measuring electrodes (both pH

and ORP)

• Yellow for reference electrodes

• Blue for combined sensors with both

measuring and reference elements

in the same body

• Green for temperature sensors

The recommended procedure is to color-code

each end of the cables to match the sensors

with the color strips provided with each cable.

This provides a quick way to identify the ends

of the cables belonging to a particular sensor

when they are installed.

3-7-1. Connection cable

The coaxial cable has two connections.

• Red to measuring element

• Blue to screen (shield)

The triaxial cable has three connections, (it has

an extra white wire termination) these wires are

connected:

• Red to measuring element

• Blue to reference

• White to screen (shield)

To connect the other sensor systems, follow the

general pattern of the terminal connections as

listed below:

Terminal

Single Measurement (pH or Redox)

Combined measurement (pH and Redox)

pH

ORP

pH and ORP

pH and rH

11

Temperature 1

Temperature 1

Temperature 1

12

Temperature 2

Temperature 2

Temperature 2

13

Reference

Reference

Reference

Reference

14

Liquid Earth

Liquid Earth

ORP

ORP

15

pH

ORP

pH

pH

16

Shield of no. 15

Shield of no. 15

Shield of no. 15

Shield of no. 15

17

Shield of no. 13

Shield of no. 13

Shield of no. 13

Shield of no. 13

Note!

• Secondary value is always temperature. For combined pH and Redox measurement, pH is the

primary value and Redox the tertiary value.

• For ORP measurement, temperature is not required for automatic temperature compensation

core

shield

red

blue

Coaxial

red

blue

white

Triaxial

core

shield

red

blue

Coaxial

red

blue

white

Triaxial