Appendix 3, temperature compensation matrix – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual
Page 64

56
IM 12B07C05-01E
The matrix can be cleared before entering new
values. Next new matrix values can be entered
as described above. The EXAxt can interpolate
the matrix. During this process it will check if
the matrix is completely ascending/descend-
ing. This is necessary as otherwise the lookup
function can give two results for one tempera-
ture. If an error is found, the EXAxt will specify
the location of the error as shown in the user
interface screen above.
The backspace key should be used for delet-
ing an individual matrix value.
An empty value is shown as
Appendix 3, Temperature compensation matrix
1. A minimum number of values is required to make interpolation possible.
The highlighted values marked as
are mandatory to enter.
Tref
T1
T2
T3
T4
T5
Sol1
S1Tr
S1T1
S1T5
Sol2
Sol3
Sol4
Sol5
S5Tr
S5T1
S5T5
2. Tref (reference temperature) is defined in the Temperature Compensation menu.
If Tref is between T1 and T5 then the value of Tref needs to be entered as T2 or T3 or T4.
Tref
T1
T2
T3
T4
T5
Sol1
S1Tr
S1T1
S1T5
Sol2
Sol3
Sol4
Sol5
S5Tr
S5T1
S5T5
3. For every SxTx that is entered the following values become mandatory to enter:
SxTr, SxT1, SxT5 and Tx
Tref
T1
T2
T3
T4
T5
Sol1
S1Tr
S1T1
S1T5
Sol2
Sol3
SxTr
SxT1
SxTx
SxT5
Sol4
Sol5
S5Tr
S5T1
S5T5