Appendix 3, temperature compensation matrix – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 64

Advertising
background image

56

IM 12B07C05-01E

The matrix can be cleared before entering new

values. Next new matrix values can be entered

as described above. The EXAxt can interpolate

the matrix. During this process it will check if

the matrix is completely ascending/descend-

ing. This is necessary as otherwise the lookup

function can give two results for one tempera-

ture. If an error is found, the EXAxt will specify

the location of the error as shown in the user

interface screen above.

The backspace key should be used for delet-

ing an individual matrix value.

An empty value is shown as

Appendix 3, Temperature compensation matrix

1. A minimum number of values is required to make interpolation possible.

The highlighted values marked as

are mandatory to enter.

Tref

T1

T2

T3

T4

T5

Sol1

S1Tr

S1T1

S1T5

Sol2
Sol3
Sol4
Sol5

S5Tr

S5T1

S5T5

2. Tref (reference temperature) is defined in the Temperature Compensation menu.

If Tref is between T1 and T5 then the value of Tref needs to be entered as T2 or T3 or T4.

Tref

T1

T2

T3

T4

T5

Sol1

S1Tr

S1T1

S1T5

Sol2
Sol3
Sol4
Sol5

S5Tr

S5T1

S5T5

3. For every SxTx that is entered the following values become mandatory to enter:

SxTr, SxT1, SxT5 and Tx

Tref

T1

T2

T3

T4

T5

Sol1

S1Tr

S1T1

S1T5

Sol2
Sol3

SxTr

SxT1

SxTx

SxT5

Sol4
Sol5

S5Tr

S5T1

S5T5

Advertising