5

IM 12B07C05-01E

3

INSTALLATION

AND
WIRING

3-1. Installation and dimensions

3-1-1. Installation site

The EXAxt 450 converter is weatherproof and

can be installed inside or outside. It should,

however, be installed as close as possible to

the sensor to avoid long cable runs between

sensor and converter. In any case, the cable

length should not exceed 50 metres (162 feet).

Select an installation site where:

• Mechanical vibrations and shocks are

negligible

• No relay/power switches are in the direct

environment

• Access is possible to the cable glands

(see figure 3-1)

• The converter is not mounted in direct

sunlight or severe weather conditions

• Maintenance procedures are possible

(avoiding corrosive environments)

The ambient temperature and humidity of

the installation environment must be within

the limits of the instrument specifications.

(See chapter 2).

3-1-2. Mounting methods

Refer to figures 3-2 and 3-3. Note that the

EXAxt converter has universal mounting capa-

bilities:

• Panel mounting using optional brackets

• Surface mounting on a plate (using bolts

from the back)

• Wall mounting on a bracket (for example,

on a solid wall)

• Pipe mounting using a bracket on a

horizontal or vertical pipe

Size nominal 50A

3. INSTALLATION AND WIRING

Figure 3-1. Housing dimensions and layout of glands

220(8.66")

184(7.2")

72(2.8")

A

D

E

F

B

C

A : For output signal

B : For contact input

C : For sensor cable

D : For contact output (S1 and S2)

E : For contact output (S3 and S4)

F : For power supply

Adapter

49

(1.93

"

)

G1/2 screw (/AFTG),

1/2 NPT screw (/ANSI)

M20 screw (/AM20)

Approx.

55(2.2

"

)

144(5.67")

144(5.67")

20

(0.79")

20

(0.79")

121.5(4.78")

24.5(1")

Adapter for Coduit Work (optional)

(option code : / AFTG, / ANSI, / AM20)

M20 cable gland

(When shipped, not installed)

Grounding terminal (M4 screw)

Hood (Option code: /H5, some cutout on the left side cover)

Unit: mm (inch)