Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual
Page 37

29
IM 12B07C05-01E
5
MENU
STRUCTURE
COMMISSIONING
mA2 similar structure to mA1
Menu
Parameter Default
Range
values
min. max.
mA1 (control)
Expire time
0.0 sec.
0 sec.
1800 sec.
mA1 (output)
Damping time
0.0 sec.
0 sec.
3600 sec.
mA1 (simulate)
Simulation perc.
50 %
0 %
100 %
P(ID)-control mA1
Setpoint
7 pH
-inf
+inf
P(ID)-control mA2
Setpoint
25ºC/ºF
-inf
+inf
P(ID)-control mA1
Range
1.00 pH
-inf
+inf
P(ID)-control mA2
Range
10ºC/ºF
-inf
+inf
P-control mA1
Manual Reset
0 %
0 %
100 %
PI(D)-control mA1
I-time
3600 sec.
1 sec.
3600 sec.
P(I)D-control mA1
D-time
0 sec.
0 sec.
60 sec.
Linear mA1
0% Value
0 pH
-inf
+inf
Linear mA2
0ºC/ºF
-inf
+inf
Linear mA1
100% value
14 pH
-inf
+inf
Linear mA2
100ºC/ºF
-inf
+inf
Table
Table mA1
see appendix 1
-2 pH
16 pH