6. wiring the sensor system, 6-1. impedance measurement jumper settings – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 20

Advertising
background image

12

IM 12B07C05-01E

Table 3-1. Impedance measuring jumpers

Jumper Settings

Application & Sensor Connections

16

13 17

15

Normal pH sensors (including FU20).

Glass sensor on Input 1.

Reference sensor on Input 2.

Special electrodes using 2 glass sensors.

(e.g. Pfaudler)

ORP (Redox measurement).

Metal sensor on Input 1.

Normal reference on Input 2.

ORP (pH compensated) or rH measurement

Metal sensor on Input 1.

pH glass (as reference) on Input 2.

63

66

65

62

61

22

21

11

12

14

IN

Imp.

LOW

IN

13

17

15

16

Imp.

LOW

Figure 3-9a. Jumper placement for low

impedance setting

Figure 3-9b. Jumper holders in cover

3-6. Wiring the sensor system

3-6-1. Impedance measurement jumper settings

Impedance measurement is a powerful

diagnostic tool. In order to perform impedance

measurements it is important to have a good

jumper setting. The table and figure below will

guide you to make the right setting.

Note! It is important to decide first which

application and which settings are

appropriate for the installation. This

decision is best done before the

jumpers are installed, because the

cables will rest on top of the jumpers in

their installed positions.

Figure 3-9a. shows the jumper positions related

to the types of measurement stated in Table

3-1.

For Low impedance the Low should be shorted

by a jumper. See drawing below.

When shipped, two jumpers are placed in a

plastic bag and supplied with the product. Typi-

cal setting for pH measurement, 13 is shorted

to become a low impedance input. Unused

jumpers should be stored in jumper holders in

the cover, as shown in Figure 3-9b.