1-2. mounting methods – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 16

Advertising
background image

8

IM 12B07C05-01E

High voltage section

Contact

(S1, S2)

output

cables

mA

cables

Contact

(S3, S4)

output

cables

Sensor

Cables

Input

contact

Power

cable

Suitable for cables with an outside diameter between 6 - 12 mm (0.24 - 0.47”)

Figure 3-5a. Cable glands diagram

3-2-2. Cables, Terminals, glands and conduit

adapter

PH450G-A(D)-A

The PH450 is supplied with terminals suitable

for the connection of finished wires in the size

range of 0.13 to 2.5 sq.mm. ( 26 to 14 AWG).

The cable glands supplied will form a tight seal

on cables with an outside diameter of 6 to 12

mm (0.24 to 0.47 inches). Unused cable entry

holes must be sealed with cable glands includ-

ing the close up plugs supplied.

PH450G-A(D)-U

The PH450 is supplied with terminals suitable

for the connection of finished wires in the size

range of 14- 26 AWG. The cable entry holes

are sealed with FM certified plugs. Prior to

cable entry the plugs can be removed with

allen key size 3/8” The cable conduit fittings

can be mounted in the holes of the housing as

required. The cable glands supplied with the

unit will give a tight seal on cables with outside

diameter of 0.24 to 0.47 inches.

Gland nut

O-ring

Gland

Close up plug

Contents:

6 X Gland M20

6 X Close up plug

6 X Gland nut M20

6 X O-ring 17.12 X 2.62 EPDM 70° sh.

NOTE: The glands must be installed properly

to meet IP66 and NEMA 4X rating.

Use close up plug for unused glands.

NOTE: Moisturize O-ring before assembling.

Figure 3-5b. How to install cable glands