3-13. hart device revision, 3-14. logbook, 4. information function – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 29: 5. setup-calibration & commissioning, 6. secondary-primary value display switch

Advertising
background image

21

IM 12B07C05-01E

4

OPERATION

OF
EXAxt

PH450G

4-3-11. HART ID = a part of the HART device ID

(descriptor)

4-3-12. Software revision = the revision level of

the software in the instrument

TROUBLE SHOOTING

If you contact the local sales/ service organiza-

tion the serial number and software revision is

necessary information. Without that information

it is impossible to help you. It is also very useful

to report all the information that you find on the

zoom-in display.

4-3-13. HART Device revision

Sometimes the firmware of a device is

updated in a way that the communication file

(HART DD) need revision too. In this case the

revision level is increased by one. The revision

level of the HART DD must match the revision

level of the Firmware. The revision level is

expressed by the first two characters of the

filename.

The following files should be used when the

HART Device revision level is 2.

(0201.aot, 0201.fms, 0201.imp, 0201.sym)

4-3-14. Logbook

The EXAxt contains several logbooks to store

history information on events, changed settings

and calibrations. The logbooks have been

categorized to simplify the retrieval of this

information.

Calibration will give information of previous

calibrations. This logbook is useful as one now

can

1) Monitor the sensor performance over time.

2) Monitor the sensor(s) lifetime.

Sensor will give all history information on

parameter settings concerning the sensor(s).

The events logged in this logbook are user

definable. This is done in

Commissioning >>

Configure Logbook >> Sensor Logbook.

Predictive maintenance. If the sensor diag-

nostics of the EXAxt are enabled, the diagnos-

tics are saved into this logbook.

For the EXAxt PH450G, the reference imped-

ance (measured between the Liquid earth and

the reference electrode) is stored every hour.

This information can be used for (predictive)

maintenance schedules as the impedance is a

measure of fouling and the sensor should be

kept clean for best results.

Settings will give all history information on pa-

rameter settings concerning the analog outputs

(mA1/mA2) and contact (S1…S4). This logbook

is useful to trace back differences in perform-

ance due to changed settings. The events

logged in this logbook are user definable. This

is done in

Commissioning >> Configure

Logbook >> Settings Logbook – mA and/or

Settings Logbook – contact

mA1/mA2 shows all (dynamic) events concern-

ing the analog outputs

S1/S2/S3/S4 shows all (dynamic) events con-

cerning the contacts.

Each HMI screen can contain up to 5 events.

As the logbook can contain 50, one can access

previous events by selecting events page 1 to

10.

4-4. Information function

In this field an information sign

, a warning

sign

or a fail sign

can appear. Pushing

this button, the user gets detailed information

about the status of the sensor or the instrument

if applicable.

See troubleshooting (chapter 8) for further

details.

4-5. Setup-calibration & commissioning

By pressing the setup

key, you get access

to the operating system of the converter based

on menus and submenus.

Browse through the list using the

key till

you find the required menu and press the

key to enter this menu.

It is also possible to press on the

or

symbol found beside the menu item.

4-6. Secondary-primary value display switch

Pressing on this text block automatically

switches the secondary value to the main

display (Large font size).

25.0