8. contact output setput – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 38

Advertising
background image

30

IM 12B07C05-01E

Expire time

If the output is over 100% for longer than the

expire time, the output will return to 0%.

Damping time

The response to a step input change reaches

approximately 90 percent of its final value

within the damping time.

100%

0%

set

point

process

value

Direct

100%

0%

set

point

process

value

Reverse

range

range

manual

reset

manual

reset

Figure 5-2. Direct/Reverse action

5-8. Contact output setup

S1/S2/S3/S4

Each Switch (contact) can have the following

functions.

1. Control : A selection of P- PI- or PID control

2. Alarm : Low or high value Limits monitoring

3. Hold : A hold contact is energised when

the instrument is in HOLD

4. Wash : See section 6-8

5. Fail

: S4 is set as fail-safe contact. 6.

6. Simulate: To test the operation of the con-

tact, simulate can be used. The

contact can be switched on or off

or a percentage of duty cycle can

be entered (DC period time)

7. Off

: Switch is not used.

power on

power on

power down normal opened contact

activated

S1, S2, S3

S4

Above table shows contact output status between common to NO.

Configure hold

Hold is the procedure to set the outputs to

a known state when going into commission-

ing. Du ring commissioning HOLD is always

enabled, out puts will have a fixed or last value.

During ca libra tion the same HOLD function

applies. For ca libra tion, it is up to the user if

HOLD is enabled or not.

Lifetime contacts

One should note that the lifetime of the con-

tacts is limited (10

6

) When these contacts are

used for control (pulse frequency or duty cycle

with small interval times) the lifetime of these

contact should be observed. On/Off control is

preferred over Pulse/duty cycle.

Setpoint

Hys.

pH

off

on

off

Delay time

Delay time

t (sec)

100

50

0

ton

toff

50%

50%

% controller output

Range

Duty cycle

ton > 0.1 sec

Duty cycle

toff > 0.1 sec

Duty cycle

100

50

0

% controller output

Range

0.3 s

Maximum pulse frequency

50% pulse frequency

No pulses

0.3 s

Figure 5-3. Alarm contact (on/off control)

Figure 5-4. Duty cycle control

Figure 5-5. Pulse frequency control