Appendix 1, Buffer tables

NIST (IEC 60746-2)/DIN 19266

°C

0 5 10 15 20 25 30 35 38 40 45 50 55 60 70 80 90 95

1,68 pH 1.668 1.670 1.672 1.675 1.679 1.683 1.688 1.691 1.694 1.700 1.707 1.715 1.723 1.743 1.766 1.792 1.806
4,01 pH 4.003 3.999 3.998 3.999 4.002 4.008 4.015 4.024 4.030 4.035 4.047 4.060 4.075 4.091 4.126 4.164 4.205 4.227
6,87 pH 6.984 6.951 6.923 6.900 6.881 6.865 6.853 6.844 6.840 6.838 6.834 6.833 6.834 6.836 6.845 6.859 6.877 6.886
9,18 pH 9.464 9.395 9.332 9.276 9.225 9.180 9.139 9.102 9.081 9.068 9.038 9.011 8.985 8.962 8.921 8.885 8.850 8.833

DIN 19267 (German buffers) so called: technical buffer solutions

°C

0 10 20 25 30 40 50 60 70 80 90

4,65

pH

DIN 4.670 4.660 4.650 4.650 4.650 4.660 4.680 4.700 4.720 4.750 4.790

6,79

pH

DIN 6.890 6.840 6.800 6.790 6.780 6.760 6.760 6.760 6.760 6.780 6.800

9,23

pH

DIN 9.480 9.370 9.270 9.230 9.180 9.090 9.000 8.920 8.880 8.850 8.820

US technical buffers

°C

0 5 10 15 20 25 30 35 40 45 50 55 60

4.0 pH US 4.000

3.998

3.997 3.998 4.001 4.005 4.001 4.018 4.027 4.038 4.050 4.064 4.080

7.0 pH US

7.120

7.090

7.060 7.040 7.020 7.000 6.990 6.980 6.988 6.978 6.970 6.890 6.980

10.0 pH US 10.317 10.245 10.179 10.118 10.062 10.012 9.966 9.926 9.889 9.856 9.828 9.828 9.828

FREE PROGRAMMABLE (Default settings based on rounded NIST values).

°C

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80

buffer

4 4.00 4.00 4.00 4.00 4.00 4.01 4.02 4.02 4.04 4.05 4.06 4.08 4.09 4.11 4.13 4.15 4.16

buffer

7 6.98 6.95 6.92 6.90 6.88 6.87 6.85 6.84 6.84 6.83 6.83 6.83 6.84 6.84 6.85 6.85 6.86

buffer

9 9.46 9.40 9.33 9.28 9.23 9.18 9.14 9.10 9.07 9.04 9.01 8.99 8.96 8.94 8.92 8.90 8.89

The freely programmable table is populated with a basic set of data to provide a start for the

user configuration. This table is intended for the user to be able to choose his buffer solutions to

suit his own preference. The data concerning the pH temperature characteristic will need to be

obtained from the supplier of the buffers.

Note: Yokogawa recommend the use of NIST (primary buffer standards) rather than buffers which

have been adjusted by the addition of acid or alkaline materials to the buffer composition. In this

way the customer gets a recognized standard, as well as the best buffer capacity (the ability to

resist pH change with contamination).

Defaults for matrix temperature compensation

Tref T1 T2 T3 T4 T5

25 ºC

5.0 ºC

25.0 ºC

45.0 ºC

65.0 ºC

85.0 ºC

Solution 1

6.40 pH

6.42 pH

6.40 pH

6.34 pH

6.23 pH

6.11 pH

Solution 2

7.00 pH

7.38 pH

7.00 pH

6.70 pH

6.45 pH

6.25 pH

Solution 3

7.30 pH

7.94 pH

7.30 pH

6.86 pH

6.54 pH

6.31 pH

Solution 4

7.60 pH

8.31 pH

7.60 pH

7.06 pH

6.67 pH

6.40 pH

Solution 5

9.00 pH

9.74 pH

9.00 pH

8.40 pH

7.91 pH

7.51 pH

Defaults for mA-output table

%

0 5

10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100

pH 0.0 0.7 1.4 2.1 2.8 3.5 4.2 4.9 5.6 6.3 7.0 7.7 8.4 9.1 9.8 10.5 11.2 11.9 12.6 13.3 14.0