17

IM 12B07C05-01E

3

INSTALLATION

AND
WIRING

3-7-3. Sensor cable connections using junc-

tion box (BA10) and extension cable

(WF10)

Where a convenient installation is not possible

using the standard cables between sensors

and converter, a junction box and extension ca-

ble may be used. The Yokogawa BA10 junction

box and the WF10 extension cable should be

used. These items are manufactured to a very

high standard and are necessary to ensure that

the specifications of the system can be met.

The total cable length should not exceed 60

metres (e.g. 5 m fixed cable and 55 m exten-

sion cable). In case of systems using dual high

impedance sensors (e.g. Pfaudler 18), then the

cable length is restricted to 20 metres (fixed

cable only, no extension with WF10).

14 Overall Screen

11

12

12

13

14

14

16

15

13

14

14

16

15

17

11

17

11 Red

12 Blue

15 Core 16 Screen
White Co-axial cable

13 Core 17 Screen
Brown Co-axial Cable

17 (screen)

14 (overall screen)

12 (blue)

11 (red)

13 (core)

16 (screen)

15 (core)

Co-axial cable

(white)

Co-axial cable

(brown)

11

12

13

14

15

16

17

11

12

13

14

15

16

17

BA10

WF10

EXA pH TRANSMITTER

/ CONVERTER

Figure 3-12. Connection of WF10 extension cable and BA10 junction box