Measuring mode wash mode recovery mode – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 50

Advertising
background image

42

IM 12B07C05-01E

Termination of a wash cycle

The user can decide to terminate the current

wash cycle. This is done in the main screen (all

other screens are deactivated) by pressing the

wash flag (once or twice). The wash cycle has

two time intervals (T

W

and T

R

) and depending

on the moment of pressing the “wash” flag the

current interval is ended (see fig. 6-1)

Note! Recovery time is intended to let the

sensor system recover to “Normal”

process conditions.

Measuring

Mode

Wash

Mode

Recovery

Mode

W

ash time

T

W

ended

W

ash flag pressed

Recovery time

T

R

ended

W

ash flag pressed

WASH

HOLD

WASH

HOLD

Interval time ended

Figure 6-1. Wash cycle

Manual wash

When enabled in the “generic wash settings”,

wash cycles can be activated manually via

the user interface:

Calibration/Wash >> Start

manual wash cycle

or via an input contact (if

provided for.)

Continuous wash during sample/hold meas-

urement

Some processes scale the electrodes in such a

way that the electrodes need continuous wash

in order to keep on performing well. In this

configuration, a “recovery time” and a “interval

time” interrupt the continuous wash (cleaning).

The continuous wash cycle is started the

moment “Continuous wash” is enabled. First

with “measuring mode” which has the duration

of “wash time”, followed by the “wash mode”

which has the duration of “interval time”.

The “interval time” and “wash time” are

reversed!

The wash cycle is terminated the same way

as described above; pressing the wash flag in

the main display (once or twice). When the

cycle is terminated “Continuous wash” should

be re-enabled to start the wash cycle.

Note! When this setup is chosen special care

should be taken in reference to scaling

when the wash cycle is terminated.

Note! In this configuration manual wash is

not applicable.

Diagnostics

The response time is a good diagnostic tool

to see the condition of the electrode system.

During the recovery time the response is moni-

tored and an error is generated when the “half

time value” was not reached within 1/3 of the

recovery time.

pH

7.0

4.5

2.0

GOOD SENSOR

AGED SENSOR

T

W

T

R

1

/

3

T

R

time

Figure 6-2. Chemical cleaning of sensors

Input Contact

The input contact is always enabled when an

output contact is configured as wash. The input

contact can be used to enable a wash after the

detection of an Impedance High error.

Advertising