9. QUALITY INSPECTION

Quality

Inspection

Standards

PH450G

pH and ORP Converter

QIS 12B07C05-01E

1st Edition: Jul. 2007

1. Scope

This inspection standard applies to the PH450G pH and ORP Converter.

2.

Inspection Items
2.1 Insulation resistance test

*2.2 Dielectric strength test

*2.3 Sensor signal input test

2.4 Temperature indication check

2.5 Current output test
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.

3.

Inspection Methods, Standards and Conditions

z

Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for

at least 5 minutes before conducting the tests. For the connections for the insulation

resistance and dielectric strength tests, follow the instructions in Sections 3.1 and 3.2.

3.1 Insulation Resistance Test

Apply 500 V DC between the terminals shown below. The insulation resistance must be

100 MΩ or greater.

(1) Between the power supply terminals shorted together (1 and 2) and the protective

earth terminal ( )

(2) Between the contact output terminals shorted together (32, 33, 42, 43, 52, 53, 72 and

73) and the protective earth terminal ( )

(3) Between the current output terminals shorted (62) and the protective earth terminal

( )

3.2 Dielectric Strength Test

(1) Apply 1390 V AC, an AC voltage of substantially sinusoidal waveform with a frequency

of 50 Hz or 60 Hz, between the power supply terminals shorted together (1 and 2) and
the protective earth terminal ( ), for at least 2 seconds. The insulation must withstand

this voltage. (The sensed current should be 10 mA.)

(2) Apply 1390 V AC, an AC voltage of substantially sinusoidal waveform with a frequency

of 50 Hz or 60 Hz, between the contact output terminals shorted together (32, 33, 42,
43, 52, 53, 72 and 73) and the protective earth terminal ( ), for at least 2 seconds.

The insulation must withstand this voltage. (The sensed current should be 10 mA.)

(3) Apply 500 V AC, an AC voltage of substantially sinusoidal waveform with a frequency

of 50 Hz or 60 Hz, between the current output terminals shorted (62) and the
protective earth terminal ( ), for at least 2 seconds. The insulation must withstand

this voltage. (The sensed current should be 10 mA.)

3.3 Sensor Signal Input Test

Connect the testing circuit as shown in Figure 1 and set the equipment as follows:

Decade resistance box (temperature simulation input):

1097.3 [Ω]

The power supply voltage should be set in accordance with the specifications of the

converter.