IM 12B07C05-01E

3

INSTALLATION

AND
WIRING

3-3. Wiring the power supply

3-3-1. General precautions

Make sure the power supply is switched off.

Also, make sure that the power supply is cor-

rect for the specifications of the EXAxt and that

the supply agrees with the voltage specified on

the textplate.

1. Install an external switch or circuit breaker

to the power supply of the converter.

2. Use an external switch or circuit breaker

rated 5A and conforming to IEC 60947-1

or IEC 60947-3.

3. It is recommended that the external switch

or circuit breaker be installed in the same

room as the converter.

4. The external switch or circuit breaker

should be installed within reach of the

operator and identified with marking as a

power supply switch to the converter.

5. Power lines such as power cables and

contact outputs should be fixed securely

onto a wall or construction using cable

racks, conduit tubing, nylon bands or

other appropriate ways. Accidental

removal from terminals by pulling may

result in electric shock.

Local health and safety regulations may require

an external circuit breaker to be installed. The

instrument is protected internally by a fuse. The

fuse rating is dependent on the supply to the

instrument. The 250 VAC fuses should be of

the “time-lag” type, conforming to IEC127.

DANGER

S1

S2

S4

S3

FRONT GLANDS

REAR GLANDS

Sensor

output

signals

HART

Contact

output

Contact

output

Power

Contact input

mA1

mA2

Figure 3-6. System configuration

Adapter for conduit work

When protect the cable with a conduit, replace the M20 cable gland with a cable gland of optional

conduit adapter, and set the adapter shown as Figure 3-5c.

Adapter

49

(1.93

"

)

G1/2 screw (/AFTG), 1/2 NPT screw (/ANSI)
M20 screw (/AM20)

Approx.

55(2.2

"

)

Packing

Unit: mm(inch)

Nut

Figure 3-5c. Conduit adapter