App index, Em starts/stops the e-mail transmission function, Cu manually recovers modbus – Yokogawa Button Operated MV2000 User Manual

Page 100: Bv enters characters, Ke key operation command, Ir resets the relative timer, Ma resets a match timer, Nr sets the trend update rate

Advertising
background image

4-27

Commands

M-4662

1

2

3

4

5

6

7

App

Index

EM

Starts/stops the e-mail

transmission function

Syntax

EM p1<terminator>
p1

Operation type

0

Start

1

Stop

Example

Start the e-mail transmission function.
EM0

Description To use the e-mail transmission function,

you must set the Ethernet interface, e-mail
addresses, and contents to be transmitted.

CU

Manually recovers Modbus

Syntax

CU p1<terminator>
p1

Communication type

0

Modbus client (Ethernet)

1

Modbus master (serial)

BV

Enters characters

Syntax

BV p1,p2<terminator>
p1

0

p2

Character string (up to 100 characters)

Example

Enter user123.
BV0,user123

KE

Key operation command

Syntax

KE p1<terminator>
p1

Key type

F1–F7

Soft keys 1 to 7

ESC

ESC key

MENU

MENU key

FUNC

FUNC key

START

START/STOP key

HISTORY

HISTORY key

USER

USER key

T/DIV

T/DIV key

0–9

Number 0 to 9 keys

MINUS

Number minus key

DOT

Number decimal key

DISP

DISP/ENTER key

UP

Up arrow key

DOWN

Down arrow key

RIGHT

Right arrow key

LEFT

Left arrow key

Example

Press the DISP/ENTER key.
KEDISP

Description Operates in the same way as the keys on the

MV. For a key sequence, transmit the commands
in the same order that you would press the keys
on the MV.

YC

Clears measured/computed data

and initializes setup data

Syntax

YC p1<terminator>
p1

Type of data to be cleared

0

Clear measured/computed data
and initialize setup data

2

Clear measured/computed data

IR

Resets the relative timer

Syntax

IR p1<terminator>
p1

Timer number to be reset

0

All timers

1–4

Timer number 1 to 4

MA

Resets a match timer

Syntax

MA p1
P1

Match timer number to be reset (1 to 4)

Example

Reset match timer number 3.
MA3

Description • An error will occur if the timer operation of the

match timer is set to repeat.

• Nothing will occur if the single match timer has

not expired.

NR

Sets the trend update rate

Syntax

NR p1,p2<terminator>
p1

1

p2

Trend update rate [/div] (15S, 30S, 1MIN,

2MIN, 5MIN, 10MIN, 15MIN, 20MIN, 30MIN,
1H, 2H, 4H, 10H)

Query

NR[ p1[,p2]]?

Example

Set the trend update rate to 5 min/division.
NR1,5MIN

Description • You cannot set the trend update rate if the

memory data type is event + display.

• The logging number is fixed to 1.
• Set the data update rate and auto save rate

first with the SW command and then set the
trend update rate with the NR command.

4.5 Setting Commands (Control)