Tb sets the bar graph display, Sb sets the bar graph of a channel, Tn sets the scale – Yokogawa Button Operated MV2000 User Manual

Page 89: Sc sets a channel display color, Ta sets an alarm point mark, Tg sets a color scale band

Advertising
background image

4-16

4.4 Setting Commands (Setting)

M-4662

SE

Sets the trend graph line width

and the number of grids

Syntax

SE p1,p2<terminator>
p1

Trend line width (1 to 3) [dot]

p2

Number of grids (4 to 12, AUTO)

Query

SE?

Example

Set the trend waveform line width to 1 dot and
the number of grids to 10.
SE1,10

TB

Sets the bar graph display

Syntax

TB p1<terminator>
p1

Bar graph display direction

HORIZONTAL
VERTICAL

Query

TB?

Example

Display the bar graph horizontally.
TBHORIZONTAL

SB

Sets the bar graph of a channel

Syntax

SB p1,p2,p3<terminator>
p1

Measurement, computation, or external

input channel number

p2

Base position of the bar graph display

NORMAL

Normal (lower limit)

CENTER
LOWER

Lower limit

UPPER

Upper limit

p3

Number of scale divisions (4 to 12)

Query

SB[ p1]?

Example

Set the number of scale divisions of the channel
002 bar graph to 5, and display the bar graph
from the span lower limit (scale lower limit if
scaling is enabled).
SB002,NORMAL,5

Description You can specify a computation channel (or make

a query) on models with the /M1 math option.
You can specify an external input channel (or
make a query) on models with the /MC1 external
input channel option.

TN

Sets the scale

Syntax

TN p1,p2,p3<terminator>
p1

Measurement, computation, or external

input channel number

p2

Display position (OFF, 1 to 10)

p3

Number of divisions (4 to 12, C10)

Query

TN[ p1]?

Example

Set the channel 003 scale position to 2 and the
number of divisions to 10.
TN003,2,10

SV

Sets the moving average of a

measurement channel

Syntax

SV p1,p2,p3<terminator>
p1

Measurement channel number

p2

Moving average ON/OFF

p3

Number of samples for computing the

moving average (2 to 400) [times]

Query

SV[ p1]?

Example

Set the number of samples for computing the
moving average on channel 002 to 12.
SV002,12

SC

Sets a channel display color

Syntax

SC p1,p2<terminator>
p1

Measurement, computation, or external

input channel number

p2

Display color (see SL (sets a trip line))

Query

SC[ p1]?

Example

Set the channel 002 display color to blue.
SC002,BLUE

Description You can specify a computation channel (or make

a query) on models with the /M1 math option.
You can specify an external input channel (or
make a query) on models with the /MC1 external
input channel option.

TA

Sets an alarm point mark

Syntax

TA p1,p2,p3,p4,p5,p6,p7<terminator>
p1

Measurement, computation, or external

input channel number

p2

Mark type

ALARM

Alarm mark

FIXED

Fixed mark

p3

Scale board display ON/OFF

p4

Alarm level 1 color (see SL (sets a trip line))

p5

Alarm level 2 color (see SL (sets a trip line))

p6

Alarm level 3 color (see SL (sets a trip line))

p7

Alarm level 4 color (see SL (sets a trip line))

Query

TA[ p1]?

Example

Display alarm marks on the channel 004 scale.
TA004,ALARM,ON

TG

Sets a color scale band

Syntax

TG p1,p2,p3,p4,p5<terminator>
p1

Measurement, computation, or external

input channel number

p2

Area (OFF, IN, OUT)

p3

Display color (see SL (sets a trip line))

p4

Display position lower limit

p5

Display position upper limit

Query

TG[ p1]?

Example

Set the color scale band range to –1.0000 to
0.5000 V (2 V range) on channel 005, and set
the display color to green.
TG005,IN,GREEN,-10000,5000

Advertising
This manual is related to the following products: