Bh sets a batch text field, Eh sets a calibration correction, Bd sets an alarm delay – Yokogawa Button Operated MV2000 User Manual

Page 93: Sm sets the custom menu

Advertising
background image

4-20

4.4 Setting Commands (Setting)

M-4662

BH

Sets a batch text field

Syntax

BH p1,p2,p3,p4<terminator>
p1

1

p2

Field number (1 to 8)

p3

Field title (up to 20 characters)

p4

Field characters (up to 30 characters)Field

title (up to 30 characters)

Query

BH[ p1,[ p2]]?

Example

Set the title to “OPERATOR” and the text to
“DAQSTATION” for field number 2.
BH1,2,OPERATOR,DAQSTATION

Description For the characters that can be used, see

appendix 1.

EH

Sets a calibration correction

If p2 is set to BEGIN

Syntax

EH p1,p2,p3<terminator>
p1

Measurement channel number

p2

Type of operation (BEGIN)

p3

Number of break points in the calibration

segment (OFF, 2 to 16)
OFF

Calibration OFF

2

to 16 Number of break points

If p2 is set to SET

Syntax

EH p1,p2,p3,p4,p5<terminator>
p1

Measurement channel number

p2

Type of operation (SET)

p3

Break point designation (1 to 16)

p4

True value of the specified break point

p5

Measured value of the specified break

point

Description • Set parameter p1 according to the table in

section 4.3.

• The selectable range of p4 and p5 varies

depending on the current set range.

• If set to scale range, the selectable range of

p4 and p5 is –30000 to 30000.

• Set true value p4 so that the value increases

as break point p3 increases.

If p2 is set to END

Syntax

EH p1,p2<terminator>
p1

Measurement channel number

p2

Type of operation (END)

Example

Example in which three break points are
specified on CH2
EH002,BEGIN,3
EH002,SET,1,0,1
EH002,SET,2,50,49
EH002,SET,3,100,101
EH002,END

Description • First, execute this command with the type of

operation set to BEGIN to specify the number
of break points.

• Set the values for the specified number of the

break points using the SET operation.

• Execute this command with the type of

operation set to END to finalize the settings.

• The EH2? command transmits the CH2

settings.

• The output example is as shown in the

example above.

• You cannot execute this command while the

MV is computing.

BD

Sets an alarm delay

Syntax

BD p1,p2<terminator>
p1

Measurement, computation, or external

input channel number

p2

Alarm delay (1 to 3600) [s]

Query

BD[ p1]?

Example

Set the channel 001 alarm delay to 120 s.
BD001,120

Description Set parameter p1 according to the table in

section 4.3.

SM

Sets the custom menu

To set the main menu

Syntax

SM p1,p2,p3,p4,p5,p6,p7,p8,p9
<terminator>
p1

Type (DISP_MAIN)

p2

to p9

Menu items to be displayed

The menu items are displayed in the
specified order.

Only the specified menu items are
displayed.
TREND
DIGITAL
BAR
OVERVIEW
INFORMATION
TRENDHISTORY
LOG
4PANEL
ESC
EXPAND
SEPARATOR

Example

Set the first menu item to TREND and the
second menu item to TRENDHISTORY.
SMDISP_MAIN,TREND,TRENDHISTORY,

Description • If you omit parameters p2 and subsequent

parameters, all menus items will be hidden.

• A command error will occur if you specify the

same menu item multiple times.

• You can specify up to three separators. If you

specify more than three, an error will occur.

• You cannot omit parameters by using

delimiters (, ,).

• 4PANEL is available only on the RD-MV2000.
• If you set the first menu item to SEPARATOR,

it will be ignored.