3. Display Description and Operation Procedure

15

3.5 MODE

Tap [MODE] to display the mode selection window (Authentication required).

*User needs to have authorization to change
measurement mode.

Choose a mode from the below selection and tap it;

“SINGLE”, “CONTINUOUS”, “INTERVAL”,
“REPEAT”, “STATS” or “STANDARD”.


Note that the instrument is capable of 180 hours of
continuous operation in every measurement mode.




Mode Description

[SINGLE MODE]

Perform a measurement once within the pre-set time frame.
Example: If the setting is as follows in “CONFIG” menu (Refer to 3.7.1 SINGLE):

INTERVAL :

N/A

SAMPLE T.

: 00:00:15

CYCLES

: N/A

A 15 second sampling is performed.

Measurement Flow
















* Delay Time: Set Delay Time in START DELAY window referring to 3.10.3 OPTION.

Sampling Time

(15sec)

Start

Measurement

Finish

Measurement

Wait Time

Pump ON

Delay Time

10 sec

Wait Time