3. Display Description and Operation Procedure

37

3.6.5 Saving Data

If AUTO SAVE is not checked on the CONFIG setting window, data will not be saved in the file after measuring.
If you try to change MODE or start another measurement when AUTO SAVE is not checked, you will get the
message shown below.


- Tap [YES] to save data.
- Tap [NO] to discard data.







If the electricity goes out during a measurement, data will be treated as below.

[SINGLE] and [CONTINUOUS] Mode

Data during a measurement will be discarded when the power goes OFF.

[INTERVAL] and [REPEAT] Mode

Data taken before electricity goes out is remained in the instrument.
When restarting the instrument, the message shown below will be displayed.


- Tap [YES] to save data.
- The calculation data will be created using data remained.
- Tap [NO] to discard data.






[STATS] and [STANDARD] Mode
Data taken before electricity went out is remained in the instrument. However, the data of the location point where
a measurement was being performed when the power went out is not saved.
The measurement is in a temporary state of suspension.

-> Refer to 3.6.6 In Case that Measurement is Interrupted.