6 start, 1 single, continuous, interval, repeat – KANOMAX 3900 Portable Particle Counter User Manual

Page 26

Advertising
background image

3. Display Description and Operation Procedure

19

3.6 START

3.6.1 SINGLE, CONTINUOUS, INTERVAL, REPEAT

Tap [START] on the main screen to start a
measurement in the setting configured in 3.5
MODE
and 3.7 CONFIG.

z If “START DELAY” is NOT checked in

“CONFIG” setting:

Upon tapping [START], the pump will begin
running. After the pump has run for 10 seconds to
stabilize, the measurement will begin


z If only “START DELAY” is checked in

“CONFIG” setting:

Upon tapping [START], the Delay Time
configured begins. During the last 10 seconds of
that Delay Time the pump will run to stabilize,
after which the measurement will begin.

z If “START DELAY” and “Set time” are

checked in “CONFIG” setting:

Upon tapping [START], “Input START TIME”
dialog box will be displayed. Enter the time when
you want to start measuring and tap [OK]. During
the last 10 seconds of the [Time left] time the
pump will start running to stabilize, after which
the measurement will begin.
Tap [CANCEL] to go back to the menu screen.
If you fail to press [OK] after entering “START
TIME”, 10 minutes will be added to the “START
TIME” automatically when the entered START
TIME comes and will continue to wait to be
activated by pressing [OK].

.



<Input START TIME dialog box>

Advertising