3. Display Description and Operation Procedure

46

3.7.6 STANDARD

STANDARD Measurement Setting
Depending on the STANDARD setting configured in
3.10.3 OPTION, the setting items for CLASS and
items in EXTENSION tab varies.

In the BASIC tab;

Configure the setting of CLASS, SCHEDULE,
ROOM, LOCATION, CYCLES and SAMPLE T.





CLASS:

Set the cleanliness level of the cleanroom to be measured. After configuring this
setting, the appropriate particle size for the configured CLASS can be selected.

SCHEDULE: You can load a MAP file. If you select the file, the MAP function will be

activated and the MAP will be displayed.

ROOM:

Input the size of the room to be measured.

LOCATION: Input the number of measurement point.
CYCLES:

Input the measurement cycles.

SAMPLE T.: Input the sampling time.
The numeric value shown next to the each entry field is the recommended value based on CLASS, ROOM and AIR
DIRECTION (in EXTENTION tab). Tap [SET] to use the recommended value.

In the EXTENTION tab, configure the followings;
- ROOM STATUS

- AIR DIRECTION (only for Federal Standard)
- COMMENT

[AIR DIRECTION]
- Select unidirectional when wind direction is

constant.

- Select nonunidirectional when wind direction is

not constant.


[ROOM STATUS]
Select either as-built, as-rest or operational.
- Select as-built when the cleanroom is setting up.
- Select as-rest when the cleanroom function is off.
- Select operational when the cleanroom function is on.

[COMMENT]
Tap the entry field to display a keypad. Enter
comment and tap [Enter].

Tap [APPLY] to activate the new settings in each tab
(Authentication required).
Tap [OK] to activate all the new settings.

(Authentication required).

Tap [CANCEL] to discard the changes and close the
window.

Check “AUTO SAVE” to store data automatically.
Check “AUTO PRINT” to print data automatically.
Check “START DELAY” to use delay time prior to a