3. Display Description and Operation Procedure

22

3.6.2 STATS

- When not using MAP (Refer to 3.7 CONFIG for using a MAP.)

Tap [START] to display “CONFIRM” window.

A measurement will start based on the setting
configured in 3.5 MODE and 3.7 CONFIG.









Tap [START] to start a measurement.

z If “START DELAY” is NOT checked in

“CONFIG” setting:

Upon tapping [START], the pump will begin
running. After the pump has run for 10 seconds
to stabilize, the measurement will begin


z If only “START DELAY” is checked in

“CONFIG” setting:

Upon tapping [START], the Delay Time
configured begins. During the last 10 seconds
of that Delay Time the pump will run to
stabilize, after which the measurement will
begin.

z If “START DELAY” and “Set time” are

checked in “CONFIG” setting:

Upon tapping [START], “Input START TIME”
dialog box will be displayed. Enter the time
when you want to start measuring and tap [OK].
During the last 10 seconds of the [Time left]
time the pump will start running to stabilize,
after which the measurement will begin.
Tap [CANCEL] to go back to the menu screen.
If you fail to press [OK] after entering “START
TIME”, 10 minutes will be added to the
“START TIME” automatically when the
entered START TIME comes and will continue
to wait to be activated by pressing [OK].

<Input START TIME dialog box>