KANOMAX 3900 Portable Particle Counter User Manual
Page 25

3. Display Description and Operation Procedure
18
[REPEAT MODE]
Configure SAMPLE T and CYCLES, and repeat measuring.
Example: If the setting is as follows in “CONFIG” menu (Refer to 3.7.4 REPEAT):
INTERVAL :
N/A
SAMPLE T.
: 00:00:15
CYCLES
: 20
20 measurement cycles lasting 15 seconds each are taken.
Measurement Flow
[STATS MODE]
Configure SAMPLE T., CYCLES and SAMPLE POINT, and repeat measuring.
Measurement at each point is same as the one in REPEAT Mode.
As for the measurement condition, you can configure the settings referring to 3.7.5 STATS.
[STANDARD MODE]
You can perform a measurement which complies with 7 measurement standards as follows:
- ISO14644-1
(International Standard) Refer to ISO14644-1.
- ISO 14644-1 SEQUENTIAL SAMPLING
(International Standard) Refer to ISO14644-1.
- FEDERAL STANDARD (m)
(The US Standard) Refer to FEDERAL STANDARD 209E.
- FEDERAL STANDARD (ft)
(The US Standard)
- BRITISH STANDARD
(British Standard) Refer to BS-5295.
- EC GMP
(European Standard) Refer to EC GMP.
- GB/T 16292-1996
(Chinese Standard) Refer to GB/T 16292-1996.
Select and set one of the above measurement standards referring to STANDARD in in
3.10.3 OPTION.
As for the measurement condition, you can configure the settings referring to 3.7.6
STANDARD.
Sampling Time
(15sec)
The 2
nd
Measurement
The 20
th
Measurement
Total Time: 5min
Start
Measurement
Finish
Measurement
Wait Time
Sampling Time
(15sec)
Sampling Time
(15sec)
The 1
st
Measurement