Connecting option sensors – KANOMAX 3900 Portable Particle Counter User Manual
Page 77

5. Connecting Option Sensors
70
5. Connecting Option Sensors
Three types of sensors are available as optional extras for this instrument.
By connecting sensors to four different connectors on the rear of the unit, you can perform three different kinds of
measurement and one contact output simultaneously.
Temperature and Humidity Probe - Model 0844
Temp. Range
0 ~ 50 °C (32~122 °F)
Temp. Accuracy
±0.5 °C (at the air velocity of 0.2m/s
or more)
Humidity Range
3 ~ 98 %RH
Humidity Accuracy
±3 %RH (±5 % when the humidity
level is outside the 30-85%RH)
Dimension
φ20 × 150mm
Air Velocity Probe - Model 0843
Velocity Range
0 ~ 1 m/s (0~197FPM)
Velocity Accuracy ±0.05 m/s (10FPM)
Dimension
φ20 × 150 mm
Curl Cord: 0.2 m (Max:1.5 m)
Pressure Transmitter - Model C264 0-100Pa
Pressure Range
0 ~ 100 Pa
Pressure Accuracy
±1 %F.S.
Guaranteed Temp
5 ~ 65 °C
Output
4 ~ 20 mA
Contact Output Cable - Model 3900-03
When the alarm function is ON and the threshold is
exceeded, alarm will be turned ON.
Spec
Photocoupler
Maximum rating: 60V/400 mA
Red: + / Black: -
[T/H]
[Air Velocity]
[D.P.]
[Alarm]
* The alarm output is a single set of contacts that
activates upon any sensor threshold being
exceeded. It can be used to activate an alarm
light, etc.
* When connecting the pressure transmitter,
cable MODEL 3900-02 is required separately.
Please check with your sales representative.