Part names and functions, 1 general outline, 2 airborne particle counter – KANOMAX 3900 Portable Particle Counter User Manual

Page 8

Advertising
background image

1. Part Names and Functions

1

1. Part Names and Functions

1.1 General Outline

This Airborne Particle Counter is a light scattering laser particle counter using semiconductor laser radiation as
a light source. This counter is used to detect ultrafine airborne particle in a cleanroom or other environment
equivalent to a cleanroom. Mainly this counter is used as a monitoring measurement equipment to monitor the
cleanliness in a cleanroom.
By using option sensors, air velocity, temperature, humidity and differential pressure can be measured
simultaneously.
Each measurement value is displayed on the LCD screen as well as can be stored onto the CF card installed
inside the counter and be printed out by using the internal printer. As for the power supply, either AC power
(100-240V) or batteries can be used.

1.2 Airborne Particle Counter




FRONT

LCD Touch Panel
--- This touch panel allows you to setup and

control all instrument functions.

Paper Cover Open Button
--- Press this button to open the cover when

changing the printer roll-paper.
To change printer paper refer to 2.5.1 Printer
Paper Replacement
.

Printer
--- For printing measurement data. Set the printer

roll-paper here.

Standard Inlet
--- Particles enter the unit from here. Attach the

provided standard inlet. To use the provided
isokinetic suction probe, attach Tygon tube on
the end of the standard inlet.

To change the standard inlet, refer to 2.2
Attaching Standard Inlet
.

Power Switch
--- When the main power switch located on the

rear of the unit is ON, you can turn ON/OFF the
power using this button.

Advertising