Instruction Manual

PD 300

To Calibrate pH:

1. If necessary, press the MODE key to select

pH mode. The pH indicator appears in the
upper right hand corner of the display.

2. Rinse the probe thoroughly with de-ionised

water or a rinse solution. Do not wipe the
probe; this causes a build-up of electrostatic
charge on the glass surface.

3. Dip the probe into the calibration buffer. The

end of the probe must be completely
immersed into the sample. Stir the probe
gently to create a homogeneous sample.

NOTE

The temperature element is in the DO electrode. For

temperature compensated readings, dip the DO

electrode into the calibration buffer as well.

4. Press

CAL/MEAS

to enter pH calibration

mode. The CAL indicator will be shown. The
primary display will show the measured
reading while the smaller secondary display
will indicate the pH standard buffer solution.

5. Wait for the measured pH value to stabilise.

See Figure 5.

6. Press

ENTER

to confirm calibration. The

meter is now calibrated to the current buffer.
The lower display scrolls through the
remaining buffer options.

For multi-point calibration, go to step 7.

For one-point calibration, go to step 9.

Figure 6: pH Calibration