Thermo Fisher Scientific CyberScan PD 300 User Manual

Page 50

Advertising
background image

Instruction Manual

PD 300

P5.1 [P4.1] READY Indicator and Auto Endpoint Function

The first program lets you select:

Figure 37: To configure for
READY and Auto Endpoint

functions

“READY indicator ON”

to indicate when the

reading is stable.

“READY indicator OFF”

for faster meter response.

Auto Endpoint Function ON

. Select auto endpoint

ON to “HOLD” the reading when it is stable for more
than 5 seconds. The display automatically freezes,
and the HOLD indicator appears on the left side of
the display. Press the HOLD key to release the
display and access other functions.

From measurement mode

1.

Press SETUP key to enter SETUP mode.

2. Press

▲ or ▼ keys to scroll through subgroups until

you view parameter COF in the upper display.

3. Press the ENTER key to select parameter “rdY”

(Ready).

4. Press

▲ or ▼ keys to select the configuration you

require.

ON switches the READY indicator on.

OFF switches the READY indicator off.

• ON and HOLD together switches the auto

endpoint feature on.

5. Press the ENTER key to confirm selection and to

proceed to:

In % mode: step 4 on page 47.

In mg/L (ppm) mode: step 5 on page 46.

Or press the CAL/MEAS key twice to return to
measurement mode.

Note: Meter default is set for Ready indicator ON, and auto endpoint function off. Any

changes made to the Ready indicator/auto endpoint function will change in the pH

mode as well.

45

Advertising