Instruction Manual

PD 300

7. Rinse the electrode with deionised water or in

rinse solution, and place it in the next pH buffer.

8. Follow steps 5 to 8 for additional calibration

points. See Figure 7.

9. When calibration is complete, press

CAL/MEAS to return to pH measurement mode.

NOTE

To exit from pH calibration mode without confirming

calibration, DO NOT press ENTER in step 6. Press

CAL/MEAS instead.

If the selected buffer value is not within

±1.0 pH from the

measured pH value: the electrode and buffer icon blink
and the ERR annunciator appears in the lower left
corner of the display. See Figure 8.

Figure 7: Next point calibration for

pH 4.01

To limit the number of pH buffer values available during
calibration, see Setup Section 8.4 P3.2: Selecting

number of pH calibration points

on page 34.

Figure 8: Err message and

electrode icon will appear if

incorrect buffer is used

11