Thermo Fisher Scientific CyberScan PD 300 User Manual

Page 19

Advertising
background image

Instruction Manual

PD 300

To calibrate 0% Saturation:

MEAS

%

%

%

ATC

°C

ATC

°C

128.5

23.4

CAL

MEAS

READY

23.4

EN

TER

CA

L

ME

AS

READY

1. Rinse the probe well with deionised rinse water. Do not

touch the membrane.

2.

Press the MODE key to select the % saturation mode.

3. Place the probe in zero oxygen solution. Stir the probe

gently to homogenise the sample. Make sure that the
sample is continuously flowing past the membrane sensor.

4. Press the CAL key. The CAL indicator will appear above

the primary display. The primary display shows the current
value of the measurement and the secondary display will
show 0.0.

5. Wait for the reading to stabilise. If the READY indicator

feature is enabled, it will appear when the reading is stable.
(See page 45 for Ready indicator setup.)

6. Press the ENTER key. The meter automatically calibrates

to 0.0% air saturation and returns to Measurement mode.

See figure on right.

NOTE

The reading in the primary display in step 3 must read at 10% or

under for the calibration to work correctly. Whenever an error occurs during calibration, the

ERR indicator appears in the lower left hand corner of the display.

Figure 11: To calibrate for

0% saturation

You can offset your % DO Calibration; see page 40 for directions.