P4.0 [P3.0] (ELE): Electrode Properties

These “view only” parameters show you the electrode
properties for diagnostic purposes:

A. P4.1 [P3.1] Probe Slope: Provides an indication of the

probe’s efficiency. The value displayed is the ratio of the
theoretical value to the actual value produced by the
probe. The higher the number, the lesser output from the
probe. The ratio displays from 0.5 to 1.999% (1.0 = 100%).

B. P4.2 % Saturation Offset: (available in % saturation

mode only):

Shows the value of the % saturation offset

entered in parameter “OFS” (see page 40 for instructions).

C. P4.3 [P3.2] 100% Saturation mV value: Lets you view

the sensor’s mV output corresponding to 100% saturation.

D. P4.4 [P3.3] 0% Saturation mV Value: Lets you view the

sensor’s millivolt output corresponding to 0% saturation.

From measurement mode:

1. Press the MODE key to select the measurement mode for

the electrode properties you want to view: % or mg/L
(ppm).

2.

Press the SETUP key to enter SETUP mode.

3. Press

▲ or ▼ keys to scroll through subgroups until you

view parameter “ELE” in the upper display.

4.

Press the ENTER key. The upper display shows the probe
slope.

5. Press the ENTER key. The upper display shows the %

saturation offset. See directions for setting this offset on
page 40. NOTE: This appears only in % saturation
measurement mode. If you are in mg/L (ppm) mode, the

meter skips to step 6 on page 43.

Figure 33: iew electrode

properties from %

saturation