P4.0: Resetting to factory default settings (pH)

This program lets you reset all pH parameters to factory default settings. This clears all
calibration data and any other pH setup functions you might have changed.

The following settings will remain as you have set them:

Temperature unit of measure (°C or °F)

The temperature offset calibration value

All DO calibration data and parameters

There are also some other parameters that retain settings when reset is done. See Addendum
3 on page 68 for a table of factory default settings.

From measurement mode

Figure 27 – P4.0: Reset to factory

default values

1. Press

the

MODE

key to select pH measurement mode.

2.

Press SETUP key to enter Set Up mode.

3. Press

the

▲ or ▼ keys to scroll through subgroups until

you view parameter P4.0 in the lower display.

4.

Press the ENTER key to enter parameter P4.0.

5. Press

the

▲ or ▼ keys to toggle between NO and YES.

See figure on right.

NO retains current settings

YES resets to factory default settings.

6. Press the ENTER key to confirm selection and to return

to the measurement mode. Otherwise press CAL/MEAS
key to return to measurement mode without resetting to
factory default.

NOTE

See page 68 for a table of factory default settings.