P1.2: Salinity Adjustment Mode

NOTE

This mode appears in mg/L (ppm) measurement mode

only.

Salinity correction mode lets you correct for the
variations in oxygen solubility due to salt concentration
in the sample.

From measurement mode:

1.

Press the MODE key to select mg/L (ppm) mode.

2.

Press the SETUP key to enter SETUP mode.

3. Press

▲ or ▼ keys to scroll through subgroups

until you view parameter “dPr” in the upper display.

4. Press the ENTER key five times. The upper

display shows the salinity value and the lower
shows SAL.

5.

Press

▲ or ▼keys to enter the correct salinity

adjustment factor. The salinity adjustment factor
range is 0.0 to 50.0 ppt.

6. Press the ENTER key to confirm selection and to

move back to subgroup “dPr”. If you want to return
to measurement mode, press CAL/MEAS.

Figure 29: Input Salinity values for
automatic correction
(The values shown above are for
illustration purpose only)