Manual temperature compensation (ph only) – Thermo Fisher Scientific CyberScan PD 300 User Manual

Page 23

Advertising
background image

Instruction Manual

PD 300

Manual Temperature Compensation (pH only)

IMPORTANT: For manual compensation, you must
disconnect the temperature probe (see page 10).

1.

Switch the meter on. Press the MODE key to
select pH mode.

2. Press

the

CAL/MEAS

key to enter pH

calibration mode. The CAL indicator will
appear above the primary display.

3. While in pH calibration mode, press the

MODE

key to enter temperature calibration

mode. The primary display shows the
current temperature setting and the
secondary display shows the default value
25

°C.

4. Check the temperature of your sample using

an accurate thermometer.

5. Press the ▲ or ▼ keys to set the

temperature to the measured value from step
4.

6. Press ENTER to confirm the selected

temperature and to return to the pH
measurement mode.

See Figure 14.

Figure 14: Manual temperature

compensation

The meter will now compensate pH readings for the
manually set temperature.

NOTE

To exit this program without confirming the manual temperature compensation value,
DO NOT press ENTER in step 6. Press CAL/MEAS instead.

5.1.3