Instruction Manual

PD 300

8.6

P1.0 (dPr): Dissolved Oxygen Parameters

This sub group is available in % saturation and mg/L (ppm) mode and lets you adjust the
barometric pressure and salinity (% saturation mode only).

8.6.1

P1.1 Pressure adjustment mode

Barometric pressure is vital to correct dissolved oxygen measurements. You need to enter the
correct barometric pressure of the area you are measuring. This mode lets you perform two
functions:

Select either mm Hg or Pascal barometric pressure units.

• Adjust the barometric pressure. See Addendum 2 on page 67 for a

“Pressure vs Altitude” table.

From measurement mode:

1. Press

the

MODE

key to select DO measurement mode.

2.

Press the SETUP key to enter SETUP mode.

3. Press

the

▲ or ▼ keys to scroll through subgroups until you view parameter “dPr” in the

upper display.

4. Press the ENTER key twice. The upper display shows either Hg or PA and the lower

display shows bAr. See figure on next page.

5. Press

the

▲ or ▼ keys to toggle between mm Hg and Pascal units.

6. Press the ENTER key to confirm selection and move to the next screen. The upper

display shows the barometric pressure and the lower display shows the units selected in
Step 3.

7. Press

the

▲ or ▼ keys to adjust the barometric pressure. The pressure adjustment

range is 500 to 1499 mm Hg (66.6 to 199.9 kPa).

8. Press the ENTER key to confirm selection and move to Step 4 on page 39 [salinity

adjustment appears in mg/L (ppm) measurement mode only

]. If instead you want to

return to measurement mode, press CAL/MEAS twice.

37