Addendum 3: meter factory default settings, 16 addendum 3: meter factory default settings – Thermo Fisher Scientific CyberScan PD 300 User Manual

Page 73

Advertising
background image

Instruction Manual

PD 300

16 ADDENDUM 3: METER FACTORY DEFAULT SETTINGS

Resetting the meter to factory default settings clears all calibration data and most other setup
functions you might have changed. The following settings will remain as you have set them:

Temperature unit of measure (°C or °F)

The temperature offset calibration value.

NOTE

DO and pH data are cleared separately from each other.

To clear pH data, see Sub group P4.0 on page 36.

To clear DO data, see Sub group P7.0 on page 49.

Type Parameter

P

No.

Default

Remarks

pH parameters

CAL

View pH calibration data

P1.1

––

Calibration data for 1st buffer, pH 1.68

P1.2

––

Calibration data for 2nd buffer, pH 4.01

P1.3

––

Calibration data for 3rd buffer, pH 7.00

P1.4

––

Calibration data for 4th buffer, pH 10.01

P1.5

––

Calibration data for 5th buffer, pH 12.45

ELE

View electrode offset

P2.1

0.00 mV

No offset adjustment

View electrode slope

P2.2

100.0 %

No slope adjustment

COF

Ready indicator

P3.1

Ready On

Ready indicator on; auto endpoint off

# pH calibration points

P3.2

3

Selects number of calibration points.

Minimum 2 points.

°C or °F

P3.3

No default

°C or °F remains as selected

rSt

pH factory default

P4.0

No

Retains your current settings

DO parameter

dPr

Barometric pressure inputs.

P1.1

Hg

Hg or Pa.

760 mm Hg

Adjustable from 500 – 1499 mm Hg

Salinity adjustment

P1.2

0 ppt

Adjust from 0 – 50.00 ppt (% only)

OFS

Set % saturation offset

P2.0

––

Up to ±10% (% only)

CAL

View calibration data

P3.0 [P2.0]

––

––

ELE

View probe slope

P4.1 [P3.1]

1.000

––

View % saturation offset

P4.2 [P3.2]

0% offset

––

View mV = 100% saturation

P4.3 [P3.3]

37.0 mV

View mV = 0% saturation

P4.4 [P3.4]

0.3 mV

COF

Ready indicator

P5.1 [4.1]

Ready On

Ready indicator on; auto endpoint off

Select mg/L or ppm units

[P4.2]

mg/L

mg/L or ppm (mg/L mode only)

°C or °F

P5.2 [P4.3]

Retains settings

°C or °F remains as selected

LCd

Adjust back lit display

P6.0 [P5.0]

8

Levels 0 to 8 (brightest)

rSt

Reset to factory default

P7.0 [P6.0]

No

Yes / no

* Program numbers with “[ ]” are accessed from mg/L (ppm) mode.