Instruction Manual

PD 300

P3.2: Selecting number of pH calibration points

Program P3.2 lets you select the number of calibration
points that appear in pH calibration mode: 2, 3, 4, or 5.
The meter will automatically exit calibration mode after
you have calibrated to your selected number of points.

From measurement mode.

1. Press

the

MODE

key to select pH measurement

mode.

2.

Press SETUP key to enter Set Up mode.

3. Press

the

▲ or ▼ keys to scroll through subgroups

until you view parameter P3.0.

4.

Press the ENTER key twice to select parameter 3.2.

5. Press

the

▲ or ▼ keys to select 2, 3, 4, or 5 point

pH calibration.

6. Press the ENTER key to confirm selection and to

return to the subgroup menu. Press CAL/MEAS
key to return to measurement mode.

See Figure on right.

Figure 25 – P3.2: Select number

of pH calibration points