Dissolved Oxygen Calibration in mg/L (ppm) mode

Calibrating the meter to 100% in saturation mode will also calibrate the mg/L mode at the
value in mg/L corresponding to 100% saturation. This should produce acceptable results in
most applications.

This meter also lets you make a calibration adjustment in mg/L (ppm) mode without affecting
your % saturation calibration. To select between mg/L and ppm units, see page 46.

NOTES ON MG/L (PPM) MODE CALIBRATION

During mg/L (ppm) measurement and calibration, the meter adjusts to the
barometric pressure value that is programmed into the meter. It also adjusts to
the salinity value that is programmed into the meter.

You can change the barometric pressure value and salinity value in the mg/L
(ppm) SETUP mode (see Section 8.6 -- P1.0 (dPr): Dissolved Oxygen
Parameters) for directions).

Whenever an error occurs during calibration, the ERR indicator appears in the
lower left hand corner of the display.