Instruction Manual

PD 300

6 HOLD

FUNCTION

This feature lets you freeze the pH, dissolved oxygen,
and temperature readings for a delayed observation.
HOLD can be used any time when in MEAS mode.

To hold a measurement, press the HOLD key while in
measurement mode. “HOLD” will appear on the
display.

To release the held value, press HOLD again.
Continue to take measurements.

NOTE

The meter shuts off automatically 20 minutes after the

last key press. If the meter is shut off either

automatically or manually, the HOLD value will be lost.

NOTE

This meter has an auto endpoint feature. When this

feature is switched on, the display will automatically

“HOLD” a reading that has been stable for more than

5 seconds. The “HOLD” indicator appears. Press the

HOLD key to release the reading. To switch on or off

the auto endpoint feature, see page 45.

Figure 19: HOLD feature

(The values shown are for illustration purpose. Your meter may display different values depending on the settings & your
environmental conditions)