Instruction Manual

PD 300

P3.3 Selecting

°C or °F

This meter lets you select between

°C and °F units for

temperature readings.

From measurement mode

1. Press

the

MODE

key to select pH measurement

mode.

2.

Press SETUP key to enter Set Up mode.

3. Press the ▲ or ▼ keys to scroll through

subgroups until you view parameter P3.0.

4. Press the ENTER key three times to select

parameter 3.3.

5. Press

the

▲ or ▼ keys to toggle between

°C and

°F.

6. Press the ENTER key to confirm selection and to

return to the subgroup menu. Press the
CAL/MEAS key to return to measurement mode.

Figure 26 – P3.3: Select

temperature units

See Figure on right.

NOTE

You can switch between °C and °F in Program P5.2 of % saturation mode (or P4.3 in mg/L

mode). If you switch between °C and °F in pH mode, the meter will also switch in % saturation

or mg/L mode.